4.4.1.1 DIAG MENU – Relays
The current state of the microwave relays is displayed (0 = OFF, 1=ON)
from leftmost (Relay #16) thru rightmost (Relay #01):
16-------------------------1
0000000000000000
REL?
See Specifications of PDL-100A located in front of this manual for your
specific configuration of delay settings for your instrument.
You can specify any one relay to toggle (if ON will turn it OFF or if OFF
will trun it ON) by entering the relay number and pressing ENTER.
Note: Enter 0 to select ALL the relays to either the ON or OFF state.
4.4.1.2 DIAG MENU – Device Calibration
An internal self-test and calibration diagnostic will be performed. The
electromechanical trombone will perform a self-test, all relays will cycle
OFF, ON, and OFF. Delay will be reset to 0.00. If an error occurs during
self-test or calibration, the instrument requires servicing or repair.
4.4.1.3 DIAG MENU – Resetting the PDL-100A
The PDL-100A will reset to a known initial POWER-ON state. If network
addresses were changed or set, the new settings will take effect.
Note: To revert to factory default settings for all device addresses
and parameters, set the GPIB address switches located on the
backpanel to ZERO (all OFF) then power-cycle (turn OFF then ON)
the PDL-100A System Unit. Remember to set the GPIB address
switches back to a valid address (1 to 31) after you configure your
device settings.
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setting Delay Values
Page 20

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pdl-100a-625psPdl-100a-1.25nsPdl-100a-10.0nsPdl-100a-20.0nsPdl-100a-100.0nsPdl-100a series

Table of Contents