Page 2
Preface Thank you for purchasing electromagnetic flow meter. Please read this manual carefully before operating and using it correctly to avoid unnecessary losses caused by false operation. Note Modification of this manual’s contents will not be notified as a result of some ...
This document will help you to establish favorable operating conditions so as to make sure that you use the equipment in a safe and effective way. In addition, something of particular attention and safety measures in the document are marked by the following marks. www.supmea.com...
Only corresponding personnel who got trained and authorized is allowed to install, use, operate and maintain the equipment. This document will help you to establish favorable operating conditions so as to make sure that you use the equipment in a safe and effective way. www.supmea.com...
Otherwise, please contact manufacturer or supplier. (1) Remote type flowmeter Figure 1 Remote type flowmeter signal converter Remote type electromagnetic flowmeter sensor www.supmea.com...
Page 10
Chapter 2 Introduction Signal cable User manual (2) Compact type flowmeter (DN65, refer to type selection manual for specific parameters) Figure 2 Compact type electromagnetic flowmeter User manual www.supmea.com...
The flow of conductive fluid induces a potential proportional to the average flow velocity, thus requiring the conductivity of the measured flowing liquid to be higher than the minimum conductivity (5us/cm). The induced voltage signal is detected by two electrodes and transmitted to the www.supmea.com...
(4) Electrode: A pair of electrodes is installed on the wall of the measuring tube which is perpendicular to the magnetic line to detect the flow signal. The material of electrode can be selected according to the corrosion www.supmea.com...
50μ S/cm. Meanwhile, the conductivity measured online must prevail, for that measured offline may be relatively higher due to carbon dioxide and nitrogen dioxide contained in the air may dissolve into the medium. www.supmea.com...
Instead, you need to check whether the parameters indicated on the nameplate are correctly preset against with the actual working conditions. The following are parameters on the nameplate. Figure 7 It is strictly prohibited to open the cover with electricity www.supmea.com -11-...
(1) Leave enough space on the side. (2) Do not make the electromagnetic flowmeter subject to violent vibration. 3.4. Pipe design (1) Location The electromagnetic flowmeter shall be installed in a dry and ventilated place. A place easy for water accumulation shall be avoided. www.supmea.com -12-...
Page 18
10 times of pipe diameters (10D), and the length of straight pipe on the downstream be at least 5 times of pipe diameters (5D) www.supmea.com -13-...
Page 19
When the electromagnetic flowmeter is required to be installed underground, the pipes at both inlet and outlet ends shall be provided with support items, and a metal protection plate shall be installed above the flowmeter. ≥10D ≥5D Figure 8 ≥5D ≥ Figure 9 www.supmea.com -14-...
Preferred produce measurement errors!) location It is easy to cause a non-full tube measurement error! Figure 11 (2) Installation direction of electromagnetic flowmeter and sensor electrodes The sensor allows horizontal and vertical installation. When it’s installed www.supmea.com -15-...
Page 21
In addition, impurities will not precipitate at the bottom of the measuring tube. It shall be guaranteed that liquids flow from bottom to top to ensure that the sensor measuring tube is always filled with medium. www.supmea.com -16-...
Page 22
(6) For pipes with open discharges, the electromagnetic flowmeter shall be installed at the bottom section (lower part of the pipe). Figure 15 (7) For places where fall head of pipes is over 5 m, the air valve shall be installed on the downstream of the electromagnetic flowmeter. www.supmea.com -17-...
Page 23
It’s not allowed to install the electromagnetic flowmeter at a place where the liquid conductivity is extremely uneven. Injection of chemicals from the upstream of the meter can easily result in uneven liquid conductivity, which can cause serious www.supmea.com -18-...
(2) In general, foreign matters (such as welding slag) may exist in newly installed pipelines. Before the flowmeter is installed, wash away the debris. It not only prevents the lining from being damaged but also measurement error caused by foreign matters which pass through the measuring tube during measurement. www.supmea.com -19-...
Then cover terminal box and tighten the screws to ensure the tightness. If the protection level of the electromagnetic flowmeter is IP68 at type selection, it has been subject to water-proof sealing. www.supmea.com -20-...
Page 26
B. Conduct regular inspection. Check against the measures mentioned above and the terminal box for at least once a year. In the event of water entry into the meter (eg, after heavy rain, etc.), the meter shall be inspected immediately. www.supmea.com -21-...
When the flowmeter is installed, the inner diameter of sealing washers shall be slightly larger than that of the pipe to aviod interference to the fluids in the pipe. Flange Nut Seal washer Bolt Figure 19 www.supmea.com -22-...
The equipment must be grounded in accordance with regulations so as to protect the operator from electrical shock. Danger! In case that equipment is used in explosion danger areas, special notes are given to explosion-proof instructions for safety tips. www.supmea.com -25-...
Page 31
Compact type (caliber: 65, refer to type selection manual for specific parameters) Figure 20 Connection description Excitation output: EXT1--Sensor excitation coil positive terminal EXT1--Sensor excitation coil negative terminal Signal output: SIG1-The positive electrode sensor signal SIG2--The negative electrode sensor signal SGND--Signal ground www.supmea.com -26-...
(3) It is not allowed to connect other electrical equipment to the grounding conductor at the same time. 4.4. Power supply Danger! The equipment must be grounded in accordance with regulations so as to protect the operator from electrical shock. (1) 100-240VAC power supply Figure 21 www.supmea.com -27-...
Page 33
: Connect ground wire to the ground screw (2) 24VDC power supply Figure 22 Tips! Allowable range: 22VDC -26VDC 24+: 24VDC Power supply positive pole 24-: 24VDCPower supply negative pole : Connect ground wire to the ground screw. www.supmea.com -28-...
485A, 485B: RS485 communication output CCOM: RS485 communication ground Protocol: ModBus-RTU Pulse, frequency output and relay out Pulse output: POUT, PCOM Relay out :ALM1,ALM2 Active mode: High 24V, 5mAdrive current Output electrical isolation: photoelectric isolation, isolation voltage: > 1000VDC www.supmea.com -29-...
Page 35
Additional remarks: pulse output is OC gate output, it needs external power supply. The general counters are equipped with pull-up resistors, and the signal can be directly connected therein. Manufacturer's suggestion: use a pull-up resistor R of 2K, 0.5W, and 24V DC power supply for power supply www.supmea.com -30-...
All parameters and hardware are configured according to your order. After energization, the instrument will perform self-check for one time. Then it will immediately begin to measure and display the current values. Figure 25 Startup interface www.supmea.com -31-...
Σ: Net flow accumulation, V: current flow rate, MT: Current conductivity] (7) Operation keys: mechanical / photoelectric keys Mark Measuring Menu mode Function mode Data mode mode > Switch menu Data right categories shift Switch Switch menu Confirmation Confirm accumulative subclass data ↲ amount www.supmea.com -32-...
Except key combination, it is forbidden to put other fingers on the other photoelectric keys when operating the touch-key. Photoelectric keys are optional, please see type selection manual for details. Figure 27 www.supmea.com -33-...
Key parameters to facilitate the manufacturer and user to quickly set up the meter: Press on at the same time to enter the parameter setting interface. Enter the password. Quickly set the password: 300000(Used to modify the quick setup menu) www.supmea.com -34-...
. When the flow unit is volume, this parameter are not displayed. Density unit : g/cm Time Figure User 0-99S constant Damping coefficient of the filter, select the average selected within the time parameter as the real-time flow. Flow Figure User 0-10% resection www.supmea.com -35-...
Page 41
The peak amplitude (not shown when peak inhibition allows configuration closing ) 1-10 Flow Option Manufacturer Y, N correction permission Indicates whether start using flow nonlinear correction function. In principle, used for small flow rate (less than 0.5 m/s) linear www.supmea.com -36-...
Page 42
Note: when set the modified point, shall keep the following relationship Modified point 1 > Modified point 2 > Modified point 3 > Modified point 4 > 0 The intermediate value of correction coefficient is 1.0000, if the correction www.supmea.com -37-...
Page 43
Flow rate modified point 3, when The flow rate function is disabled, this parameter does not display. 1-17 Flow Figure Manufacturer 0.0-99.999 correction coefficient 4 Flow rate modified point 4, when The flow rate function is disabled, this parameter does not display. 1-18 Flow Figure Manufacturer 0.0-99.999 1.000 www.supmea.com -38-...
Page 44
User 0-5000 2000 frequency Set the corresponding frequency of the real-time flow upper limit ; When selecting frequency output, this parameter displays. Pulse value Option User 0.001-999.99 ( L/P) Set the cumulant that each pulse stands for ; www.supmea.com -39-...
Page 45
Touch spot outputs high and low level under no alarm state . Alarm1 output Option User allowed Allow contact 1 output of main switch, when it’s set to be N, the following parameters do not display. Allow alarm1 Option User empty pipe www.supmea.com -40-...
Page 46
Used to eliminate the alarm disturbance Upper limit elimination conditions: real-time flow is less than the upper limit alarm value minus return difference Lower limit elimination conditions: real-time flow is greater than the lower limit alarm value plus return difference www.supmea.com -41-...
Page 47
EVEN Serial communication verification mode of physical layer 2-1 4-3、3-4 2-1 4-3 1-2、4-3 1-2、 Byte swap Option User 1-2 3-4 Byte exchange sequence of physical layer serial communication Device Figure User 0-999999 000001 address HART device identification number www.supmea.com -42-...
Page 48
Measured conductivity equivalent value when the tube is full, default values can be used for general natural water. It needs to observe whether the empty pipe for special fluid is that displayed as 9-1, then record it in 9-4. Empty pipe Figure Manufacturer 0-9999 detection www.supmea.com -43-...
Page 49
For correcting the sensor’s nonlinear correction for small flow (below 0.3 m/s) For details, see sensor coefficient calibration section. Excitation Option Manufacturer 3.125Hz 、 6.25Hz mode 6.25 Hz 、 12.5 10-7 Hz 、 25 Hz Selection of excitation frequency 3.125Hz 、6.25Hz、12.5Hz、25 Hz www.supmea.com -44-...
Page 50
Source code Option Manufacturer After set to Y, the original code of the signal will be displayed on the 11-2 running screen, and this screen will display the firmware version number and product serial number at the same time www.supmea.com -45-...
Users can conduct the switch operation in the menu by pressing and button, switch among the parameter item of menu by pressing the button and meanwhile store a modified parameter value. Adjust the parameter value by pressing button. Such as flow upper limit. Figure 29 www.supmea.com -46-...
Sensor signal is greater than the AD sampling of the upper limit AD_Hi The current real-time flow rate exceeds the setting flow limit The current real-time flow rate exceeds systemAD sampling limit Rng_Hi The range scope set by user exceeds the upper limit of pulse output. Pls_Hi www.supmea.com -51-...
Pulse equivalent output is different from frequency output; pulse output will output a pulse when a pulse equivalent is accumulated enough, so the pulse output is uneven. Counter instrument should be used when measuring pulse output. Frequency meter instrument shouldn’t be used. www.supmea.com -52-...
Indicate real-time flow rate real time Indicate current instrument range Indicate real-time current value real time 7.3. Communication This instrument provides a standard RS485 communication interface, using the international standard MODBUS-RTU communication protocol that supports 04 Read Holding Registers command. www.supmea.com -53-...
Reverse flow ulong The decimal part magnifies accumulation of by 1,000 times 123 stands decimal for 0.123 Note: Float/ulong/long type data, Communication transmission is in byte order2-1-4-3; ushort type data transmission is in accordance with 2-1. www.supmea.com -54-...
(only IP68 protection of remote type sensor) Immersion depth Less than 3 meters (only IP68 protection of remote type sensor) Sensor cable Only for remote type, the standard 10m cable; other cables suggest custom no longer than 20 meters. www.supmea.com -56-...
Page 62
Operating Environment Temperature -10 ℃ - 55 ℃ for Compact-Type Flowmeter -10 ℃ - 60 ℃ for Converter of Remote-Type Flowmeter Environment -10 ℃ – 55 ℃ for Converter of Remote-Type Flowmeter Storage -40 ℃ - 65 ℃ -57- www.supmea.com...
Page 63
Output pulse width: 0.25ms ~100ms Pulse output basis Duty cycle: 50% (Pulse frequency ≥5Hz) Fmax ≤ 5000 cp/s setting 0.001L – 1m frequency Fmax ≤ 5000Hz setting 0-5000Hz passive ≤ 36VDC Outer Status output function Output as alarm passive ≤ 36VDC Outer www.supmea.com -58-...
So this table is only for reference. Users may make their ownchoices based on actual situation. You may refer to corrosion prevention manual for general media. But for media with complex compositions like mixed acid, you may need to conduct corrosion tests for materials to be selected. www.supmea.com -59-...
The flowmeter is not limited to 0~10mA(DC) or 4~20mA (DC) standard current output and 1 ~ 5kHz frequency output at the same time. Because the flowmeter ( sensor ) has the above-mentioned advantages, it has www.supmea.com -63-...
The highest temperature of the measured medium: ABS60℃ Shell protection level: It complies with the relevant provisions of IP68 of GB-08-84 standard. (10) Sensor output signal: 0.209mVp-p/1m/s. (11) The maximum transmission distance between the sensor and the converter is www.supmea.com -64-...
Faraday's law of electromagnetic induction. Terminal box: The junction box is located on the top of the sensor. The terminals in the junction box act as a connection between the sensor and the converter. www.supmea.com -65-...
Page 71
Sealing mechanism: composed of pressing screw seat, pressing nut, rubber washer and set screw made of stainless steel material. It is used to seal and insert the electromagnetic sensor so that it can withstand a certain working pressure. www.supmea.com -66-...
Need help?
Do you have a question about the SUP-LDG and is the answer not in the manual?
Questions and answers