Temperature Monitoring - AGT PSG BCR05 Installation Manual, Operating Manual, Maintenance Manual, Spare Part List

Sample gas cooler
Table of Contents

Advertisement

Installation-, Operating-, Maintenance-Manual, Spare Part List
Temperature-Monitoring and Error-Diagnosis
Temperature-Monitoring
The monitoring unit displays the measured cooling temperature of the sample gas (black hand) as well as
the defined lower and upper limits (green and red hands).
NOTICE: Temperature display! A dew point within 0°C and +10°C displayed on the control
unit is correct according to the possible working conditions (gas flow-rate, gas inlet
temperature, gas inlet dew point, ambient temperature, etc.). If the dew point goes under 0°C
or over +10°C a malfunction is reported over the potential free alarm contact.
Sample gas cooling temperature (black hand)
Reference value: +3 to +7°C
Error-Diagniosis / Checklist
Fault
Failure of sample gas cooler
Sample gas flow blocked
Condensate in gas outlet
Temperature > +10°C
Temperature < 0°C
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Cause
Interrupted power supply
Protective switch activated
Defective refrigerant circuit
Gas paths contaminated
Gas paths frozen
Gas inlet/outlet connected incorrectly
Condensate outlet blocked
Sample gas cooler overloaded
Defective refrigerant circuit
Too low ambient temperature
Defective performance regulation
Lower limit 0°C (green hand)
Upper limit +10°C (red hand)
Remedy
Re-establish the power supply
Check operating conditions,
reset protective switch manually
Contact AGT Service
Contamination may occure due to dust or
sublimate. Ensure pre-separation
Adhere to operation data
Connect gas inlet/outlet correctly
Clear condensate outlet,
check installation
Check operating conditions,
ensure proper cooling air circulation,
condensor cleaning
Contact AGT Service
Adhere to operation data
Contact AGT Service
Translation - Manual BCR05 Rev. 06/2017
BCR05
Page 17