AGT PSG BCR05 Installation Manual, Operating Manual, Maintenance Manual, Spare Part List page 10

Sample gas cooler
Table of Contents

Advertisement

Installation-, Operating-, Maintenance-Manual, Spare Part List
Assembly and Installation
Feed Lines
DANGER! Condensate may be toxic or corrosive! Please wear sufficient protective clothing
(for example safety gloves, face mask, safety goggles etc.).
CAUTION! Toxic and explosive gases! Corresponding guidelines must be adhered to while
operating with toxic and explosive gases.
NOTICE: Environment protection! Don't dispose the condensate in the environment. Don't
drain the condensate into the sewage system without treating it first. Dispose the condensate in
compliance with the local rules.
Regardless of the type of heat exchanger system (see chapter "Technical Data") always mount
an appropriate , pneumatically tight drainage (eg. peristaltic pump, float drain, condensate
tank, etc.). For each gas path and condensate outlet a separate drain must be provided.
Without suitable condensate drainage the device will not function properly.
Lay the condensate drain line outside the device with an adequately dimensioned gradient and
without counter pressure. Avoid unwanted pipe bends and level differences as well.
If the dew point in the sample gas pipeline between the outlet probe and the sample gas
conditioner inlet falls, the sample gas pipeline must be equipped with a steam trace to avoid
condensate waste. Connect heated discharge lines to the cooler with suitable thermal
decoupling.
In case of condensation in the sample line, a suitable pre-separator needs to be installed before
the cooler. We also recommend to use a pre-separator in case of measuring points at high inlet
temperatures and high inlet dew points greater than 65°C. The capacity of heat exchangers is
thus affected favorably.
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Translation - Manual BCR05 Rev. 06/2017
BCR05
Page 10

Advertisement

Table of Contents
loading

Table of Contents