Microlife MT1P21-PRO Quick Start Manual

30-second dishwasher safe thermometer

Advertisement

Available languages

Available languages

Quick Links

30-Second
Dishwasher Safe Thermometer
Model #MT1P21-PRO
This predictive clinical digital electronic thermometer is for the intermittent
measurement and monitoring of human adult and pediatric body temperatures;
orally, and under the arm.
This thermometer can be used with or without
a tip cover. To use a cover, slide onto sensor
tip until fully engaged and peel off paper.
Covers are for single use, do not reuse. If
thermometer has been in a cold environment,
allow it to warm up to room temperature for
about 15 minutes before use. Avoid storing at
temperatures above 140˚ (60˚ C).
This thermometer IS intended to be used:
• For self-measurement/monitoring of human body temperature whereas the
person being measured may be the user/operator of the device.
• Orally, or Under the Arm (axillary).
• Adults
• Pediatric
This thermometer IS NOT intended for use with:
• Neonatal
General Warnings
• Contact your physician in case of prolonged fever as this requires medical
attention, especially for young children.
• Self-measuring means monitoring, not diagnosis or treatment. Under no
circumstances should you independently alter the dosages of any drugs
prescribed by your physician.
• Do not use this device if you think it is damaged or anything appears unusual
(e.g. sporadic operation, open packaging upon purchase).
• To avoid possible injury, do not run or walk while using this device.
• Ensure that children are supervised around this medical device; some parts are
small enough to be swallowed.
• Keep batteries away from children; harmful if swallowed.
• Do not dispose of the battery or device in fire; the battery may explode.
• Do not use a metal tool to replace the battery; doing so may short circuit the terminals.
• Do not attempt to recharge the battery as they are not rechargeable.
• Only open the thermometer battery cover to replace the battery.
• Do not open/modify this device; inaccuracy and/or harmful injury may result.
• Performance of this thermometer may be degraded if used outside of its specified
operation temperature and humidity rating, or if stored outside of its specified
storage temperature and humidity rating. Avoid dust or storage in direct sunlight.
• Performance of this thermometer may be degraded if the device is exposed to
mechanical shock.
• Performance of this thermometer may be degraded if the patient's temperature is
lower than that of the ambient.
• Performance of this thermometer may be degraded if the tip is dirty or damaged;
ensure the tip is clean before using.
• Performance of this thermometer may be degraded if the probe cover is dirty or
a probe cover not intended for use with this device is used.
• Any probe covers that may be included in this device package are for single use only.
• This device contains sensitive electronics components. Avoid strong electrical or
electromagnetic fields in the direct vicinity of the device (e.g., mobile telephones,
microwave ovens). These can lead to temporary impairment of the measuring
accuracy. Move the device to another location if interference is determined.
Maintain a minimum distance of 3.3 m from such devices when using this unit.
• According to international standards, your device should be checked for
accuracy every 2 years.
How to Take a Temperature:
Press power button to turn unit on. Display should read "188.8" to confirm
normal function. An "Err" message indicates unit is not working properly. The
last recorded measurement will show for 2 seconds, followed by "- - -" and a
flashing "F" or "C", indicating that thermometer is ready for use.
Function Check
Oral Use:
• Do not drink hot or cold liquids or exercise before use.
• Place sensor tip under tongue as shown in Figure A.
• Accurate temperature reading will take approximately 30
seconds. An open mouth or improper placement could
make reading take longer.
Underarm Use:
• Wipe underarm with dry towel.
• Place sensor tip in armpit and lower arm against body as
shown in Figure B.
• Accurate temperature reading will take approximately 5
minutes with proper use.
Reading the Temperature:
For readings 99.5˚ F (37.5˚ C) and below, the thermometer
beeps 10 times to indicate temperature is within normal range.
For readings above 99.5˚ F (37.5˚ C), the thermometer emits a series of short,
rapid beats to indicate a possible high temperature.
Last Recorded
Ready to
Temperature
Measure
(Figure A)
(Figure B)
Sensor Tip
Flexible Probe
Changing from °F to °C:
With the thermometer turned off, press and hold the power button for 5
seconds. An "°F" or "°C" will appear on the display, blinking. Press the
ON/OFF button to toggle to the opposite temperature setting. After 4 seconds
your thermometer will be ready for use.
How to Clean:
Clean thermometer after every use. There are 2 cleaning options below:
Clean
1. Alcohol:
sensor tip with 70% isopropyl alcohol. Wipe dry with
sterile cotton ball.
2. Dishwasher Safe: To clean in dishwasher, place thermometer on top rack only.
Run dishwasher on normal cleaning cycle. High heat drying cycle is NOT recom-
mended.
How to Change Battery:
Replace battery when "
battery (1.5V, LR41 or equivalent). Plus (+) sign should face upward. Replace
endcap to close.
Limited Warranty:
This thermometer is warrantied for 5 years by Microlife USA Inc, against
manufacturer defects for the original purchaser only, from date of purchase.
The 5 year warranty applies to the thermometer. Batteries and case are not
covered by this warranty. There are no user serviceable parts inside.
The warranty does not apply to consequential and incidental damages, or
damage caused by batteries, improper handling, and accidents. Professional
use, not following the operating instructions, and alterations made to the
device or accessory by third parties, are also not included in this warranty.
Some states do not allow the exclusion or limitation of incidental or consequential
damages, so the above limitation or exclusion may not apply to you.
Microlife USA Inc will investigate your concern. A product as defined by this
warranty, determined to be out of specification, will be replaced and shipped
to you at no cost. A product or accessory as defined by this warranty, determined
to be within specification, will be returned to you with a report of findings, at
no cost.
Please call customer service: 1-800-568-4147 regarding any warranty concerns.
We ask that you please contact us before sending any product back in order
to better identify, and more quickly process, your concern.
This warranty gives you specific legal rights, and you may also have other
rights which vary from state to state.
Technical Specifications:
Measurement range:
Measurement Accuracy:
Self test:
Display:
Auto Off:
Smallest Unit of Display:
Memory:
Operating Temperature:
Storage temperature:
Battery:
Expected product life:
Disposal:
Reference to Standards:
IP67:
Electromagnetic Compatibility: Device fulfills the stipulations of the standard
Test Protocol (for laboratory use only):
1. Place thermometer into waterbath. The tip of the thermometer should be ap-
proximately 2cm below the surface.
2. Wait for 10 seconds, then turn the thermometer on by pressing the "On/Off"
button. The thermometer is now in standard mode.
3. When the measurement is completed, record the temperature reading.
Distributed by:
Microlife USA, Inc.
1617 Gulf to Bay Blvd., 2nd Floor
Clearwater, FL 33755
©2020 Microlife Corporation
www.microlifeusa.com
Follow instructions of use to avoid hazards.
LCD Display
" appears on display. Replace old battery with new
90.0° F to 109.9° F (32.0°C to 43.9° C)
±0.2° F (±0.1° C) between 93.2° F (34° C) to
107.6° F (42° C)
Automatic internal check of sensor circuit. If
error, «Err» is displayed.
Liquid crystal display (LCD) with four digits
With reading: after 9 minutes of inactivity
Without reading: after 1 minute of inactivity
0.1° F (0.1° C)
For storing the last measured value
50° F(10 °C) to 104° F (40° C) humidity of 15% to
95% (non-condensing)
-13° F (-25° C) to 140° F (60° C) humidity of 15%
to 95% (non-condensing)
1.5V, LR41
5 years
Batteries and electronic instruments must be
disposed of in accordance with the locally ap-
plicable regulations, not with domestic waste.
Device Standard: Device corresponds to the
requirements of the standard for electronic
thermometers ASTM E-1112, ISO 80601-2-56,
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 as
well as ISO 10993-10:2010, ISO 10993-5: 2009,
ISO 10993-12:2012
Totally protected against dust ingress,
protected against the effect of immersion
between 15cm and 1m.
IEC 60601-1-2
Questions?
We're here to help. Call toll-free
1-800-568-4147
Type BF applied part
End Cap
Power Button

Advertisement

loading

Summary of Contents for Microlife MT1P21-PRO

  • Page 1 • Do not use a metal tool to replace the battery; doing so may short circuit the terminals. Microlife USA Inc will investigate your concern. A product as defined by this • Do not attempt to recharge the battery as they are not rechargeable.
  • Page 2 Garantía Limitada: funcionamiento esporádico, el envase estaba abierto al momento de comprarlo). Su termómetro digital está garantizado por Microlife USA Inc., por cinco años, • Para evitar posibles lesiones, no correr ni caminar durante el uso de este dispositivo. por defectos de manufactura solamente para el comprador original desde la •...