IFX-F100 Operating Instructions Index Page Safety instructions ................4 Symbols used in these operating instructions ........4 Safety instructions ................5 Introduction ..................6 How to order ..................7 Installation .................... 8 Unpacking and storage ..............8 3.1.1 Unpacking ................8 3.1.2 Storage ..................
Page 3
IFX-F100 Table of Contents 4.4.4 Datalogger ................25 Commissioning ................... 25 Menu structure ................25 Diagnostics ..................31 Display settings ................31 5.3.1 Main PV ................. 31 Output settings ................31 5.4.1 Serial interface RS 232 ............31 5.4.2 Serial interface RS 485 / Modbus RTU ........31 5.4.3 HART compatible output ............
Safety instructions Symbols used in these operating instructions Danger This symbol represents an immediate hazardous situation which could result in a serious injury, death or damage to the equipment. Where this symbol is shown, do not use the equipment further unless you have fully understood the nature of the hazard and have taken the required precautions.
• Install the equipment and cabling securely and safely according to the relevant regulations. • If the product does not operate normally, please refer to the ser- vice and troubleshooting instructions, or contact Isoil Industria for help. Flow_Transmitter_IFX-F100_EN_2018_March_Rev.0...
Introduction Clamp-on The IFX-F100 is a hand-held, battery operated ultrasonic flowmeter employ- transit-time flowmeter ing clamp-on sensors for the measurement of liquids in full, enclosed pipes. Flow measurements can be undertaken without interruption of the process or interference with the integrity of the pipeline. The clamp-on sensors are attached to the outside of the pipes.
Installation Unpacking and storage Unpacking 3.1.1 Care should be taken when opening the box containing the flowmeter, any markings or warnings shown on the packaging should be observed prior to opening. The following steps should then be taken: • Unpack the flowmeter in a dry area; •...
IFX-F100 User Manual Clamp-on sensor installation The correct selection of the sensor location is crucial for achieving reliable measurements and a high accuracy. Measurement must take place on a pipe in which sound can propagate (see Acoustic Propagation) and in which a rotationally symmetrical flow profile is fully developed (see Straight Pipe Lengths).
Installation location Select an installation location following the recommendations in Table 1 and try to avoid measuring: • in the vicinity of deformations and defects of the pipe, • near welding seams, • where deposits could be building up in the pipe. For a horizontal pipe: Select a location where the transducers can be mounted on the side of the pipe, so that the sound waves emitted by the transducers propagate horizontally in the pipe.
Page 11
IFX-F100 User Manual Disturbance source: 90°-elbow Inlet Outlet L 10 D L 5 D Disturbance source: 2 x 90°-elbows in one plane Inlet Outlet L 25 D L 5 D Disturbance source: 2 x 90°-elbows in different planes Inlet Outlet L ...
Disturbance source: valve Inlet Outlet L 40 D L 10 D Disturbance source: pump Inlet L 50 D Table 2: Recommended distances from disturbance sources Pipe preparation • Clean dirt and dust from around the area of the pipework where the sensors are to be placed;...
Page 13
IFX-F100 User Manual Illustration 3: Clamp-on sensor mounting configurations and sensor spacing Transducer The transducer separation distance A is measured from the inside edges of separation distance the sensor heads as shown in illustration 3. It is automatically calculated by the flowmeter based on the parameter entries for pipe outside diameter, wall thickness, lining material and thickness, medium, process temperature, the sensor type and the selected number of signal passes.
Flowmeter installation 3.6.1 Outline dimensions The IFX-F100 is a wall mounted device and can be installed using suitable screws and wall plugs according to the following drawings. Flowmeter outline dimensions Drilling aid for all mounting Make sure that the ambient temperature is within the -10 ... 60 °C operating temperature range specified for the flowmeter unit.
IFX-F100 User Manual 3.6.2 Electrical connections Please note that in order to supply the unit with MAINS POWER, the equipment must Electrical wiring be protected by suitably sized switches and circuit breakers. 100 ... 240 V AC, 50/60 Hz 10 W 9 ...
Clamp-on sensor mounting Before the sensors can be mounted Sensor mounting • the installation location should have been determined, • a sensor mounting method should be chosen, • the flowmeter must be mechanically and electrically installed, • the sensors must be connected to the flowmeter. Depending on which sensor mounting method is being used, the clamp on sensors are either mounted on the same side of the pipe (Reflection Mode) or on opposite sides of the pipe (Diagonal Mode).
IFX-F100 User Manual 3.7.1 Sensor pipe mounting configurations Illustration 4: Sensor pipe mounting configurations 3.7.2 Acoustic coupling gel In order to obtain acoustical contact between the pipe and the sensors, apply a bead of acoustic coupling gel lengthwise down the centre of the contact area of the sensors.
The sensor separation distance is automatically calculated by the flowmeter based on the parameter entries for pipe outside diameter, wall thickness, lin- ing material and thickness, medium, process temperature, the sensor type and the selected number of signal passes. 3.7.4 Sensor mounting with tension straps Illustration 6: Metallic mounting straps •...
IFX-F100 User Manual • Ensure that the narrower side of the clip is above and inside the wider side and that the two sides of the clip do not come into con- tact while tightening, as this will prevent the strap from being cor- rectly tensioned.
Backlight on/off ESCape menu item Abort entry without saving, escape measurement mode ENTer menu item Confirm entry with saving or move through menu structure Table 3: Keyboard function 4.2.2 Display functions Menu header Display icons Display line 2 Display line 1 Date, Time Display icons Display icon...
IFX-F100 User Manual Error recorded in error log No error detected Serial output switched on Serial output switched off "L", "T" or "LT" Displays whether flow is laminar, turbulent or mixed Quick setup wizard The quick setup wizard allows for a speedy setup of the most important pa- Quick start wizard rameters in order to achieve successful measurements in the shortest pos- sible time:...
Page 22
Choose pipe material using cursor keys and pressing <ENTER>. Enter outside pipe diameter using alphanumeri- cal keys and confirm by pressing <ENTER>. Use key <UP> as character backspace clear to correct for data entry errors. If 0 is entered, an additional screen appears that allows entering the pipe circumference.
Page 23
IFX-F100 User Manual Use cursor keys to select Write Flowmeter. Confirm by pressing <ENTER>. The programmer will send setup parameters to the flowmeter. Use cursor keys to select Sensor Positioning. Confirm by pressing <ENTER>. Use cursor keys to select Start Measurement. Confirm by pressing <ENTER>.
Measurements 4.4.1 Main process value (PV) display Measurement is started using the Quick Setup Wizard. Once all the param- eters are programmed, any subsequent power-on sequences will bring up the main PV display immediately. Display screen Operation The main process value can be changed in the menu structure.
IFX-F100 User Manual 4.4.2 Diagnostic displays Diagnostic screens Display screen Operation Line 1 shows the amplifier gain. Line 2 displays the signal strength. Line 3 indicates the noise. Change to more diagnostic displays by pressing <NEXT>. 4.4.3 Totalisers The totaliser displays will only be shown when the totalisers are activated. Totalisers Display screen Operation...
Page 26
K4N,K4L,K4E,K4Ex,K4P, K0, M, Q, Special Middle (main Select from list where available ↑↓ displayed) m/s, f/s, in/s, m3/h, m3/min, m3/s, l/h, l/min, l/s, Units USgal/h, USgal/min, USgal/s, bbl/d, bbl/h, bbl/min, g/s, t/h, kg/h, kg/min, m3, l, Usgal, bbl, g, t, kg, W, kW, MW, J, kJ, MJ, Signal (dB), noise (dB), SNR (dB), C m/s (measured sound speed), CU (housing...
Page 27
IFX-F100 User Manual Off, On, Reset + (positive total), Reset – (negative total) Reset Both Start Measurement Sensor type Indication of sensor type and serial number if au- tomatically detected, otherwise select from list ↑↓ As Setup Wizard Sensor fre- SP1, only for special, unrecognised sensors quency Wedge angle...
Page 28
Damping Reduces fluctuations in the display output 1 ... 255 s Metric / Imp Use metric or imperial units for entered data In/Output Type Select from list ↑↓ Current out Source Channel 1 System Units Select from list ↑↓ Min Value Min.
Page 29
IFX-F100 User Manual Alarm – PV alarm switch Math – Calculated value alarm switch Fault – Allocated to system failures, see error re- port list On Point Value of PV at which the relay energises when in alarm mode Off Point Value of PV at which the relay de-energises when in alarm m ode Current In...
Page 30
User Identifier Example: Pump P3A 9 character string Tag No. Example: 1FT-3011 9 character string Test Installation Control system simulation: A cyclic repetition of in- creasing flow velocity across the measureable range. All configured outputs respond as if this was a measured change in flow. Yes, No Display Display screen test routine...
IFX-F100 User Manual 9600, 19200, 57600,115200 Parity Select from list ↑↓ None Even (Default) Table 5: Firmware menu structure Diagnostics Diagnostic screens, where specified, can be viewed directly during measure- ment using the programmer or through the menu structure (screen only). Display settings The main Process Value (PV) is the primary measurement data.
Wiring Setup Please refer to customer support. Operation Please refer to customer support. 5.4.3 HART compatible output The KF150 can also be configured with an optional module which responds to output commands conforming to the HART protocol. Please refer to customer support for further information.
IFX-F100 User Manual 5.4.5 Analogue voltage output 0 – 10 v Voltage outputs may be assigned to process values in the “mode” section of the output menu. The outputs can be programmed and scaled within the menu structure. Wiring Electrical Galvanically isolated from main electronics and from other I/O's.
Wiring Electrical Form A (SPDT-NO and NC) contacts characteris- Width 3...990 ms. tics U=48 V, Imax=250 mA.Galvanically isolated from main electronics and from other I/O's. Mode: Alarm, fault, totaliser (programmable). 1 Form A (SPST-NO) contacts. 1 Form A (SPST-NC) contacts. Width 3...990 ms.
IFX-F100 User Manual If a heat quantity unit is specified for the Process Value, the KF100 will prompt the user for the Specific Heat Capacity of the medium in J/g/K (for example 4.186 J/g/K for water). This may also be entered in the System\Calculation sub-menu. The In/Output menu will then allow the user to select the temperature input source;...
• This means that Isoil can only service this device if it is accompanied by a Return Authorization Number (RAN) confirming that the device is safe to handle.
Page 37
IFX-F100 User Manual Error list Error message Group Description Error handling USB INIT FAIL Hardware Internal board communi- Power on/off, otherwise cation error call customer support NO SERIAL NO. Hardware Failed to read from FRAM Call customer support NO VERSION NO. Hardware Failed to read from FRAM Call customer support PARA READ FAIL...
SENSOR COU- Application Weak sensor coupling, Recouple sensors, PLING ERR low SNR check installation, re- duce number of passes, look for other location, call customer support Table 6: Error messages Data download difficulties If difficulties are encountered downloading the logger data: - Check that the flowmeter is switched on and not in measurement ●...
IFX-F100 User Manual Specification General Measuring principle : Ultrasonic time difference correlation principle Flow velocity range : 0.01 ... 25 m/s Resolution : 0.25 mm/s Repeatability : 0.15 % of measured value ± 0.015 m/s Accuracy : ± 1 ... 3 % of measured value depending on application, ±...
Page 46
Process inputs / Process Outputs (maximum of four per instrument) Inputs Temperature : PT 100, three or four-wire circuit, measuring range - 50 ... 400 °C, resolution 0.1K, accuracy ±0.2 K Current : 0 ... 20 mA active or 4 ... 20 mA passive, U = 30 V, Ri = 50 Ohm, accuracy 0.1 % of MV Outputs Current : 0/4 ...
/dichiariamo sotto la nostra responsabilità che i seguenti prodotti: Product model/Modello: Ultrasonic flowmeter/Misuratore di portata ad ultrasuoni: Isoflux IFX-F100, IFX-P200, IFX-P210 and associated trasducers/ e i trasduttori associati are in conformity with the EEC directives /sono conformi alle seguenti direttive europee:...
Page 48
For EU Customers only - WEEE Marking. Marking of electrical and electronic equipment in accordance with Directive 2012/19/EU This symbol on the product indicates that it will not be treated as household waste. It must be handed over to the applicable take-back scheme for the recycling of electrical and electronic equipment.
Page 49
ISOIL INDUSTRIA S.p.A HEAD OFFICE SERVICE Via Fratelli Gracchi, 27 isomagservice@isoil.it 20092 Cinisello Balsamo (MI) Tel +39 02 66027.1 Fax 039 026123202 sales@isoil.it If you want to find the complete list of our distributors access at the following link: http://isoil.com/u_vendita.asp...
Need help?
Do you have a question about the Isoflux IFX-F100 and is the answer not in the manual?
Questions and answers