Setting Attenuation Levels

Because the OA 5000 automatically adjusts the attenuator to correct for
different wavelengths, it is important, for attenuator accuracy, to correctly
match the displayed wavelength with the wavelength of the signal going into
the unit.
Setting attenuation levels on the OA 5000 requires you to perform two
procedures in the following order:
1. Specify the wavelength of the signal to be attenuated.
2. Set the level of attenuation.
Refer to Figure 2-3.
REMOTE
MODE
SET REF
ADDR
WAVE-
(SET)
LENGTH
Figure 2-3: Setting Attenuation Levels
To specify the wavelength to be attenuated:
Step 1: Press the WAVELENGTH button.
The word SEt will appear in the attenuation display and the WAVE-
LENGTH button will illuminate.
DISABLE
MIN ATT
STORE 1
RECALL 1
STORE 2
RECALL 2
COARSE
FINE
This button sets
attenuation to the
minimum
value.
These knobs set the
attenuation level,
reference level,
wavelength and GPIB
address.
2---5