LSI LASTEM Heat Shield Quick User Manual
Hide thumbs Also See for Heat Shield:

Advertisement

Quick Links

Heat Shield
Quick user guide
INSTUM_01515_en
MW6175
upd. 21/12/2015

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Heat Shield and is the answer not in the manual?

Questions and answers

Summary of Contents for LSI LASTEM Heat Shield

  • Page 1 Heat Shield Quick user guide INSTUM_01515_en MW6175 upd. 21/12/2015...
  • Page 2: Instrument Description

    INSTRUMENT DESCRIPTION (1) – LCD display 4x20 chrs. (2) – Keyboard (3) – Base mounting (4) – Tripod mounting (BVA325) (5) – Reset (6) – Power supply / battery charger (7) – Anemometer input (8) – RS232 port (9) – Wet bulb temperature sensor (10) –...
  • Page 3: Switch On/Off

    . Press to confirm. Press to exit. Measurement display Heat Shield MASTER shows measurements and indexes from its sensors and, when con- nected, from SATELLITE units. moves to different group of measurements/indexes. to scroll the group to the next/previous one.
  • Page 4: Data Logging

    . Press to stop data logging. Status bar It contains some Heat Shield working information: where: a: storing in progress. nn: survey number. ss:mm: can have the following meanings: - on programmed survey, it is the time to reach the starting time of the sur- vey (s=”-“).
  • Page 5: Change Parameters

    SETUP It is possible to change some factory setup parameters. Browsing Push to go on the MAIN menu. to scroll each in formation, and to enter; to return to previous selection, it goes to Measurements when pushed from MAIN menu. Change parameters Depending by the situation it is possible to change the parameters from a selectable list (ex- ample: language) or using the arrows (example: day/time).
  • Page 6: Getting Started

    Survey setup Go to Survey setup. 1. Nr survey: Heat Shield assign a progressive number to each survey. It is possible to change (1 ÷ 99) the first assigned number during measurement. 2. Mode: Heat Shield can performs measurement from the base unit only (WBGT1 amb), from MASTER unit plus one SATELLITE (WBGT 2 amb), from MASTER unit plus two SATELLITE units (WBGT 3 amb).
  • Page 7: Installation

    HS Manager is a PC tool used to download the stored data and perform data analysis. Installation Run Setup and follow its instructions. During the first run, the program asks to install a new Heat Shield unit. 1. Connect Heat Shield MASTER to PC via RS232 To detect the PC port where the port (8) using its cable and USB adapter.
  • Page 8 EXAMPLES Fig. 1 - WBGT 1 environment measurement and index from MASTER unit only. Fig. 2 - WBGT environments: separate measurements and indexes from MASTER unit and one SATELLITE unit.
  • Page 9 Fig. 3 - Installation of the units in case of WBGT Subject. Fig. 4 - WBGT subject: measure- Fig. 5 - Usage of WBGT 1 env. ments and indexes averages from with cup anemometer. the MASTER and 2 SATELLITE units. Fig.
  • Page 10 NOTES This manual is an introduction to the use of Heat Shield. For more information refer to the INSTUM_01517 User’s guide included on the MW6501 - LSI Lastem Products DVD. Disposal Heat Shield is a highly electronic scientific device. In accordance with the standards of envi- ronmental protection and collection, LSI LASTEM advises to handle Heat Shield as waste of electrical and electronic equipment (WEEE).
  • Page 12 www.lsi-lastem.it...