12
schwa-medico
Instructions for Use - TENStem eco basic

"Ready" mode

The LCD display shows the selected program, the therapy time, the frequency, the pulse
width and the intensity.
1.
Press the
P
button to select the programs.
2.
Press the left " button: The TENStem eco basic changes into the mode "therapy" and
the intensity for channel 1 increases by 1 mA.
3.
Press the right
the intensity for channel 2 increases by 1 mA.
If electrodes that are not correctly connected to the TENStem eco basic are
Caution!
placed on the skin, a current of 9-10 mA or more causes the intensity to be reset to zero.
4.
Simultaneously press and hold the left
seconds for series A and B devices): the program select function of the
the program parameter editing function are disabled. The lock symbol
in the display. With this function you can ensure that a setting of the device that has
been selected for use can not be accidentally changed. By pressing the left
P
and the
button again for 3 seconds, the program select function of the
the program parameter editing function are enabled again.
After locking or unlocking the device, the program must first be started via a "-"
button. Only then the locking or unlocking will be stored and stays functional even
after switching the device on and off.
Only for series A and B devices:
5.
Simultaneously press and hold the left
total stimulation duration.
E
Press the
button again to return to the "Ready" mode.
Only for devices from series C upwards:
E
6.
Press the
button to enter the "Edit" mode (see also section ""Edit" mode").
Press and hold the
also section ""Date/Time" mode").
Only for devices from series D upwards:
7.
Reset all editable parameters: Press the button
editable parameters of the device to the as-delivered condition. The device confirms
the execution of the function with a long-lasting acoustic signal. The therapy time
of all programs (except program 16: 20 minutes) is set to 30 minutes. The following
default values are assigned to the parameters frequency and pulse width of the user
programs:
Program
13
14
15
16
8.
Switch the signal transmitter on and off: Press the
simultaneously. After three seconds, the current status of the signal transmitter is
shown in the display. "BEEP ON" means the signal transmitter is switched on. "BEEP
button: The TENStem eco basic changes into the mode "therapy" and
E
button for three seconds to access the mode "Date/Time" (see
Frequency Hz
Puls width µs
100
150
80
100
50
200
35
200
P
and
buttons for three seconds (two
and
E
buttons for two seconds: Display of the
for three seconds to reset all
Timer min
30
30
30
20
E
and the right
P
button and
appears
button
P
button and
buttons