How To Clean Instruments With A Fixed Connector Interface; How To Clean Instruments With An Optical Glass Plate; How To Clean Instruments With A Physical Contact Interface - Agilent Technologies 81618A User Manual

Optical head interface and optical heads
Table of Contents

Advertisement

Which Cleaning Procedure should I use ?
Cleaning Information
How to clean instruments with a fixed
connector interface
You should only clean instruments with a fixed connector interface
when it is absolutely necessary. This is because it is difficult to remove
any used alcohol or filaments from the input of the optical block.
It is important, therefore, to keep dust caps on the equipment at all
times, except when your optical device is in use.
If you do discover filaments or particles, the only way to clean a fixed
connector interface and the input of the optical block is to use
compressed air.
If there are fluids or fat in the connector, please refer the instrument
to the skilled personnel of Agilent's service team.
CA U TI O N
Only use clean, dry compressed air. Make sure that the air is free of
dust, water, and oil. If the air that you use is not clean and dry, this
can lead to filmy deposits or scratches on the surface of your
connector interface. This will degrade the performance of your
transmission system.
Never try to open the instrument and clean the optical block by
yourself, because it is easy to scratch optical components, and cause
them to become misaligned.
How to clean instruments with an optical
glass plate
Some instruments, for example, the optical heads from Agilent
Technologies have an optical glass plate to protect the sensor. Clean
this glass plate in the same way as optical lenses (see "How to clean
lenses" on page 100).
How to clean instruments with a physical
contact interface
Remove any connector interfaces from the optical output of the
instrument before you begin the cleaning procedure.
Cleaning interfaces is difficult as the core diameter of a single-mode
fiber is only about 9 µm. This generally means you cannot see streaks
or scratches on the surface. To be certain of the degree of pollution on
the surface of your interface and to check whether it has been
removed after cleaning, you need a microscope.
101
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User's Guide, Second Edition