5.1 Function ..........13 Table of contents 5.2 Programs ..........13 1 Introduction ........3 5.3 Maximum pressure (caring mode)..13 1.1 General information ......3 5.4 Panel lock ..........13 1.2 Intended use ......... 3 5.5 Comfort settings ........13 1.3 Intended user ........
Page 3
10.1 General ........26 10.2 Between users ......26 10.3 Replacing the air filter ....26 11 Troubleshooting ........27 12 Technical specification .......28 12.1 Standards ........28 12.2 Symbolkey ......... 29 13 Other information......30 13.1 Recommended lifetime of the product ..........30 13.2 Disassembly and recycling ..
Page 4
The product must only be installed and used for its intended purpose according to the instructions in this manual and/ or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts (e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other documentation from Care of Sweden.
Page 5
Contact an authorized service technician for examination and repair. Never connect anything other than the Care of Sweden supplied power supply to the control unit power cable connector.
Introduction Mattress Recommended user weight These air mattress systems may be used as an CuroCell® OP10 Up to 160 kg aid to prevent and treat pressure ulcers/pressure injuries. CuroCell® CX10 Up to 200 kg CuroCell® M4 is a control unit where the CuroCell®...
Assembly and installation Check that no parts are damaged. If damage is found, contact Care of Sweden or your local distributor before using the product. Do not use sharp objects when unpacking as it might damage the product. When using an overlay mattress (CuroCell®...
Transport function Common operations If the user needs to be moved in bed, either: Following operations apply to CuroCell® M4, • Unplug the CPR connection, close the lid, CuroCell® A4 and CuroCell® IQ regardless of place the CPR connection at the end of the which mattress that is used.
® CuroCell Operation CuroCell® M4 Following operations apply only to CuroCell® M4 regardless of which mattress that is used. Operation CuroCell® M4 (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2.
Page 10
® CuroCell Button Function Mute the information signal Panel lock Pack & Go®. Function for deflating the system Alternating mode Pulsating mode CLP mode (Constant low pressure) Maximum pressure (caring mode) User weight settings Information signal Incorrect connection of the air connector (CPR) Seating function Cycle time settings (10, 15, 20, 25 minutes).
® CuroCell Function At start-up, the inner pressure of the mattress must be set manually based on the weight and Maximum pressure (caring height of the user. The mattress system mode) maintains the pre-set inner pressure regardless With this function, the entire of movement and position changes.
® CuroCell Sitting positioning in bed Hand check (function control) See 3.1 Hand check is performed to ensure that the CPR (Cardiopulmonary resuscitation) mattress system works properly and to ensure that the weight setting is correct. Hand check See 3.2 should be performed regularly;...
® CuroCell 4.15 Table of notifications Information about each notification is shown in the notification table: • Notifications from 1-7 are both audible and visual. The error code will be displayed until the error has been rectified. If the mute button is pressed the audible warning will crease for a period of 5 minutes and will return until the error has been rectified.
® CuroCell Operation CuroCell® A4 Following operations apply only to CuroCell A4 regardless of which mattress that is used. Operation CuroCell® A4 (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2.
Page 15
® CuroCell Button Function Mute the information signal Panel lock Pack & Go®. Function for deflating the system Alternating mode Pulsating mode CLP mode (Constant low pressure) Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR) Cycle time settings (10, 15, 20, 25 minutes).
® CuroCell 3. Pulsating mode. The cycle period Function can be changed according to the user The mattress system independently and needs and requirements. The longer without manual adjustment controls the inner the cycle period, the slower the alternations. pressure of the mattress according to the We recommend a basic setting of 10 minutes.
® CuroCell Press the Pack&Go® button and hold it Note! down for 2 seconds • Make sure that the mattress system is filled, which is shown by a green light The Pack&Go® diode will flash during deflation. from the diode, before performing a The mattress will empty of air and be ready to hand check.
Page 18
® CuroCell Notification Description and troubleshooting High temperature. Valves and compressors are turned off. If the control unit is in direct sunlight, relocate it. Default settings are not completed. Incorrect input voltage. Make sure that the correct power supply is used, otherwise contact the technical support. Low pressure.
® CuroCell Operation CuroCell® IQ Following operations apply only to CuroCell® IQ regardless of which mattress that is used. Operation CuroCell® IQ (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2.
Page 20
® CuroCell Button Function Mute the information signal Pack & Go®. Function for deflating the system Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR) Check-symbol. System is ready to use Notification diodes...
® CuroCell When using an overlay mattress system Function (CuroCell® OP10 or CuroCell® CX10) The mattress system independently and 1a. Open the cover and identify a cell with less without manual adjustment controls the inner air at the user’s sacrum. Insert a hand, with the pressure of the mattress according to the palm facing up, between the overlay mattress weight, length, and position of the user.
® CuroCell Pack&Go® The product may be packed as follows: Ensure that no-one is lying on the mattress. On the control panel, press the lock/unlock button. Press the Pack&Go® button and hold it down for 2 seconds The Pack&Go® diode will flash during deflation. The mattress will empty of air and be ready to be simply folded together within 20 minutes.
® CuroCell 6.12 Table of notifications Information about each notification is shown in the notification table: • Notifications from 1-10 are both audible and visual. The error code will be displayed until the error has been rectified. If the mute button is pressed the audible warning will crease for a period of 5 minutes and will return until the error has been rectified.
Product description Control unit (M4/A4/IQ) CuroCell® M4 is a manual air mattress system used as an aid to prevent and treat pressure ulcers/pressure injuries. The inner pressure of the mattress must be set manually based on the weight of the user. Perform a hand check to make sure the pressure is correct, see section 4.6. The mattress system maintains the pre-set inner pressure regardless of movement and position changes.
MATTRESS Reuse and cleaning 1. Disconnect the tube connector from the The product is reusable. Follow the instructions control unit and remove the air from the below for cleaning and reconditioning. It is mattress. recommended to clean the mattress between 2.
We recommend that the control unit will be regularly serviced and inspected to maintain functionality and performance. Service and maintenance must always be performed by Care of Sweden or one of its authorized technicians. Only use spare parts approved by Care of Sweden. For more information, see the service manual for CuroCell®...
11 Troubleshooting If the problem keeps occurring, please contact Care of Sweden or your local distributor. Problem Solution The control unit Check that the power supply has been connected to the mains supply. does not start Check that the LED on the power supply is showing green.
12 Technical specification Note: Care of Sweden reserves the right to modify the product specification at any time. CONTROL UNIT SPECIFICATION CUROCELL® M4/A4/IQ Model CuroCell® M4, A4, IQ Input voltage 100-240 V / 50-60 Hz / 0,6 A Output voltage...
12.2 Symbolkey Symbols to convey medical device information CE-marked in accordance with Medical Device Manufacturer Regulation (EU) 2017/745 Medical Device Distributor Symbols for traceability and product information Item number Type BF Class II Equipment (double SN-number insulated). Indicated on the power supply.
• If it is assessed that the product is or could be contaminated (e.g., used by users with a known bloodborne infection), the product must be handled in accordance with the healthcare 13.3 Returns Contact Care of Sweden or your local distributor before returning the product.