Preparation Of The Pipette For Testing - Vitlab Micropipette Manual

Hide thumbs Also See for Micropipette:

Advertisement

2. Preparation of the pipette for testing

2.1 Instrument identification
Read instrument type and nominal capacity
Read serial number (located behind tip ejection
button)
Read customers identification, if present
2.2 Minimum configuration for VITLAB micropipette
Have the operating manual ready
VITLAB micropipette
®
Pipette tips
2.3 Cleaning
Clean pipette shaft
Clean the exterior sufficiently
Any liquid residues inside the device?
For multichannel pipettes:
V- resp. O-rings damaged?
4
► Enter in test record
Enter number in test record
Enter identification in test record
®
Use only manufacturerer's original parts
► Use only appropriate tips.
Optimum results can only be obtained with
VITLAB pipette tips.
Wipe off with a soft cloth, so that no media
residues remain
Use a damp cloth (Water or diluted soap
solution)
Disassemble and clean the instrument (see
operating manual)
No media residues! V- and O-rings can be
replaced (see operating manual)
®