Figure 2-12, Setting the Enhanced Monitoring System Configuration Switch

Configuring The 1261B 2-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Omit this procedure if your mainframe is not equipped with the
Enhanced Monitoring System (EMS).
Perform this procedure only if factory default setting
is incompatible with your application. See Figure 2-12
for default setting.
Ensure that J400 AC Input is disconnected before
removing or replacing the System Monitoring (EMS)
module. Possible damage to the module could result.
1. Remove the EMS module. See Removal
2. Configure
3. Replace the EMS module. See Removal
1261B User Manual
NOTE
WARNING
and Replacement in Chapter 7.
Position "1" enables changes.
SW1
located
component side of the EMS module.
and Replacement in Chapter 7