The MIN, MAX, and
DEF Parameters
Querying
Parameter Setting
SCPI Command Reference 5-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Each of these has a long form and a short form. Thus
MEAS:VOLTAGE2?
MEAS:VOLT2?
are equivalent.
Omitting the keyword suffix selects the same capability as
specifying the suffix of 1. This means that:
MEAS:VOLTage?
MEAS:VOLTage1?
are equivalent.
Substitute MINimum, MAXimum, or DEFault in place of a
numeric parameter for most commands. For example, the
following command:
:VOLTAGE3:RANGE:UPPER MIN
sets the upper voltage range for the -2V power supply monitor to
its minimum programmable value. The command:
:VOLTAGE3:RANGE:UPPER DEF
sets the upper voltage range for the -2V power supply monitor to
its default, power-on value.
The MINimum, MAXimum, and DEFault keywords are accepted
for
all
SENSE:<Attribute>:RANGE
VXI:CONF:MON:TRIG:DELAY command.
Query the current value of most parameters by adding a
question mark ( ? ) to the command. For example, the following
command sets the trigger delay to 0.5 seconds:
VXI:CONF:MON:TRIG:DELAY 0.5
Query the trigger delay by executing:
VXI:CONF:MON:TRIG:DELAY?