ECHNICAL ATA AND PECIFICATIONS 7.0 R ............................... 30 EPLACEMENT UMBERS 7.1 ENMET part numbers for sensors and replacement parts: ......................30 7.2 ENMET part numbers for Calibration equipment:......................... 30 A: CO C ..............................31 PPENDIX HARACTERISTICS B: G ............................32...
Page 3
GSM-60 ENMET List of Figures Figure 1: External Features of the GSM-60 ........................5 Figure 2: GSM-60 Interior Features ........................... 6 Figure 3: GSM-60 Mounting Dimensions ......................... 7 Figure 2A: Relay, Input and Output Terminals ........................8 Figure 4: GSM-60 Operational Display ........................... 11 Figure 5: GSM-60 Operation Menu Flow Chart ......................
1.2 Check Order Check, the contents of the shipment against the purchase order. Verify that the GSM-60 is received as ordered. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
GSM-60 ENMET 2.0 Instrument Features 2.1 Exterior Features The exterior of the instrument is shown in Figure 1. The exterior features are as follows: Feature Description Enclosure An engineered thermoplastic box, approximately 10x8x6, with a clear hinged front cover. Input Port The entrance for the air sample and calibration gas.
Mounting Flanges 2 places Sample Output Port Figure 1: External Features of the GSM-60 : Typical gas reading & alarm locations, depending on instrument configuration, alarms & readings may be in alternate locations Manual Revision Date – June 15, 2017 P a g e Manual Part No.
Removes contaminate from air sample line. Sensor Terminals Sensor/Transmitter connectors 24V 4-20mA Input J16, J18, J19 Figure 2: GSM-60 Interior Features Manual Revision Date – June 15, 2017 P a g e Manual Part No. – 80003-600 Rev. 1 – December 15, 2016...
3.1 Mounting of Instrument The GSM-60 should be located near the air to be monitored. Sampling lines should be no more than 50 feet long. It is recommended that Teflon® (PTFE) lined tubing be used. Quick disconnect fittings are supplied for use with 1/8” ID tubing.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the switch. If alarm persists longer than 30 minutes, call ENMET AUDIO DEFEAT ALARM ACKNOWLEDGE customer service personnel.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
These relay contacts can be used to operate auxiliary alarms or other functions. It is recommended that power for auxiliary equipment be supplied from an independent power source, separate from the GSM-60. Place a hole in the enclosure for a wire exit, and use appropriate cable fittings.
4.1 Normal Operation Condition With the GSM-60 installed as described in Section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display is as shown in Figure 4 Display, for the sensor(s) installed in the GSM-60. The red alarm and fault LEDs are not lit.
Differential Setting The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value. Example: If the alarm point is Λ 10 and the differential is 2, the GSM- 60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.
5000 Function Enter Maint Menu Enter Password See Maintenance Menu Diagram Figure 5: GSM-60 Operation Menu Flow Chart Manual Revision Date – June 15, 2017 P a g e | 13 Manual Part No. – 80003-600 Rev. 1 – December 15, 2016...
4.8 Hydrocarbon Sensor Response If a Hydrocarbon (HC) sensor is supplied with the GSM-60 instrument, it designed to detect hydrocarbon gases and vapors with an ionization potential (IP) of 10.6 eV or less. Hydrocarbons with an IP of greater than 10.6 eV will NOT be detected. Please see Appendix B for a list of common gases and vapors and their respective IP rating.
: Never spray a cleaning solution on the surfaces of the GSM-60 devices. AUTION Clean the exterior of the GSM-60 enclosures with a mild soap solution on a clean, damp cloth. Do not soak the cloth with solution so that moisture drips onto, or lingers on, external surfaces.
ELECT Exit maint menu See Section 5.3.9 for instructions on how to exit Maintenance Menus. Figure 6: GSM-60 Maintenance Menu Flow Chart. Manual Revision Date – June 15, 2017 P a g e | 16 Manual Part No.
Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the...
5.3.1A Low Cal/Zero Cal Adjust A Low-Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.
GSM-60 ENMET 5.3.1B High Cal/Span Gas Adjust A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed. Press the switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to SELECT "HI Cal xx"...
Example: If the alarm set point is Λ 10 and the differential is 2, the GSM-60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.
Page 22
GSM-60 ENMET Example: Set Alarms Flow Chart Displays are examples of Alarms Λ - Indicates alarm triggered on increasing value of reading O = Press Option v - Indicates alarm triggered on decreasing value of reading S = Press Select...
GSM-60 ENMET 5.3.4 Set Alarm Delay The alarms may be set to delay by 1 second increments, up to 255 seconds. Alarm delays are factory set to 5 seconds. To change an alarm delay, you must enter the maintenance menu. Press the switch until “Enter Maint Menu”...
GSM-60 ENMET 5.3.5 Relay Configuration To change a relay configuration, you must enter the maintenance menu. Press the switch until “Enter Maint Menu” is OPTION displayed then press switch for the Enter Password menu. Enter the valid password as described below.
S = Press Select 5.3.7 Set Output Span Range To change 4-20 mA output range. This range is set at the factory and should not be changed, contact ENMET for information. Press the switch to continue to next section of maintenance menu.
When all four digits of the new password have been selected, "Set New Password" is displayed. Record the new password; without it, the maintenance menu cannot be reentered once you exit the Maintenance Menu. If the password is lost, call ENMET customer service personnel. From the "Password XXXX" position, ...
Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the...
Page 28
: Instruments equipped with an oxygen sensor programmed with a range of 0 to 15% by volume require 99.999 % Nitrogen (ENMET part number 03295-100) be used for the Low Cal/Zero Cal setting and 5% by volume Oxygen (ENMET part number 03296-050) be used for HI Cal/Span setting To continue to next section, Press the switch.
5.4.3A Low Cal/Zero Cal Adjust A Low Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.
GSM-60 ENMET 5.4.3B High Cal/Span Gas Adjust A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed. Press the switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to SELECT "Hi Cal xx"...
GSM-60 ENMET 7.0 Replacement Part Numbers 7.1 ENMET part numbers for sensors and replacement parts: Part number Description 03053-000 Sensor, CO2 67025-1114 Sensor, Oxygen 03015-014 Sensor, MOS 812 03016-014 Sensor, MOS 813 03028-XXX Sensor, PID, 10.6 eV Contact ENMET 67025-XXXX...
If a CO sensor is installed, the GSM-60 has two preset alarm set points, at 10 ppm and 20 ppm CO, which are adjustable, but cannot be set below 5 ppm or above 50 ppm.
GSM-60 ENMET Appendix B: Gas Ionization Potentials Chemical Name IP (eV) Chemical Name IP (eV) p-Bromotoluene 8.67 2-Amino pyridine 8.00 p-tert-Butyltoluene 8.28 Acetaldehyde 10.21 s-Butyl amine 8.70 Acetamide 9.77 s-Butyl benzene 8.68 Acetic acid 10.69 sec-Butyl acetate 9.91 Acetic anhydride 10.00...
Page 34
GSM-60 ENMET Chemical Name IP (eV) Chemical Name IP (eV) (D continued) Decaborane 9.88 2-Furaldehyde 9.21 Diazomethane 9.00 Fluorine 15.70 Diborane 12.00 Fluorobenzene 9.20 Dibromochloromethane 10.59 Formaldehyde 10.87 Dibromodifluoromethane 11.07 Formamide 10.25 Dibromomethane 10.49 Formic acid 11.05 Dibutylamine 7.69 Freon 11 (trichlorofluoromethane) 11.77...
Page 35
GSM-60 ENMET Chemical Name IP (eV) Chemical Name IP (eV) (P continued) 2-Methyl furan 8.39 2-Pentanone 9.38 2-Methyl napthalene 7.96 2-Picoline 9.02 1-Methyl napthalene 7.96 3-Picoline 9.02 2-Methyl propene 9.23 4-Picoline 9.04 2-Methyl-1-butene 9.12 n-Propyl nitrate 11.07 2-Methylpentane 10.12 Pentaborane 10.40...
Page 36
GSM-60 ENMET Chemical Name IP (eV) Chemical Name IP (eV) (T continued) Trimethyl amine 7.82 Water 12.59 Tripropyl amine 7.23 2,4-Xylidine 7.65 o-Vinyl toluene 8.20 m-Xylene 8.56 Valeraldehyde 9.82 o-Xylene 8.56 Valeric acid 10.12 p-Xylene 8.45 Vinyl acetate 9.19 Vinyl bromide 9.80...
Page 37
ENMET, LLC for damages arising out of or in connection with the use or repair or performance of the product.
Page 38
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired.
Page 39
Mailing/Shipping Address: ENMET 680 Fairfield Court Ann Arbor, MI 48108 Phone: 734.761.1270 repair@enmet.com Fax: 734.761.3220 Service Request Form Product Name or Number: Product Serial Number: Describe Problem or Needed Service: Warranty Claim? ☐ Yes ☐ No CUSTOMER INFORMATION Billing Address:...
Need help?
Do you have a question about the GSM-60 and is the answer not in the manual?
Questions and answers