EATURES OF THE 3.0 I EX-5130-IS ......................... 3 NSTALLATION OF THE 3.1 Mounting the EX-5130-IS Enclosure ..........................3 3.2 Wiring the EX-5130-IS to a Control Unit .......................... 4 4.0 O EX-5130-IS ..........................5 PERATION OF THE 4.1 Start up ....................................5 Typical Start Up .......................................
The ENMET EX-5130-IS sensor/transmitters (S/T) is a three wire 4-20mA S/T for the detection of toxic gas. The EX-5130-IS is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX-5130-IS sensor/transmitter has been designed to be used in a Class I, Div. 1, Groups B, C, D, classified areas.
Three alarm points are preprogrammed into the EX-5130-IS sensor/transmitters. At each alarm point, an LED on the front panel is activated. These internal alarm settings are independent of the 4-20mA output alarm values that can be set at a controller.
, 4-20 mA S/T for the detection of toxic gas. The S/T is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX- 5130-IS sensor/transmitter has been designed to be used in a Class I, Div. 1, Groups B, C, D, classified areas.
Figure 3: Terminal Positions EX-5130-IS Sensor/Transmitter When wiring is complete, reassemble the EX-5130-IS. Use caution when installing the overlay so as not to damage the magnetic switches. Put the cover back on the S/T Do Not apply power to the S/T without the cover in place.
ENMET Corporation 4.0 Operation of the EX-5130-IS It is best to have the EX-5130-IS transmitters powered up and operational for 24 hours before applying calibration or test gas to them. When the EX-5130-IS transmitter is first powered up, it goes through a series of momentary screens, which identify the instrument model number, serial number and software revision.
4.2 Normal Display Mode When the EX-5130-IS is installed as described in section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display is measurement of the target gas detected by the EX-5130-IS. The red alarm and fault LEDs are not lit.
Hold the magnet over M switch for 2 – 5 seconds to enter the Maintenance Menu The Power/Fault LED will flash Green – Red to indicate the EX-5130-IS is in Maintenance Mode To exit the maintenance Menu and return to the Normal...
Another cylinder is used to provide the Span reference point for calibration. Depending on the instrument calibration, the Span gas may be the same gas that the instrument is calibrated to display, or it may be another gas, which ENMET has found to have a similar response.
5.2.1 Zero Adjust A ZERO function should be performed only when the EX-5130-IS sensor/transmitter is exposed to fresh air. If the air at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. Attach the cylinder to the calibration adapter.
Page 12
ELECT If the Zero signal is within Preset Specs the EX-5130-IS will display Cal OK, See Section 5.2.1 If the Zero signal is not within Preset Specs the EX-5130-IS will display Bad ZERO Bad ZERO : Some software revisions require the S...
Section 6.0 of this manual. If you do not know the proper part number for your sensor, be sure to have the EX-5130- IS serial number available when contacting your Distributor or ENMET Corporation Technical Support. To replace a sensor, it is not necessary to open the transmitter housing.
Calibration Gas, Consult ENMET Distributor or ENMET Corp. 7.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc...
Page 15
Fax 734.761.3220 Returning an Instrument for Repair ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.