Cleaning And Disinfecting - Welch Allyn KleenSpec 78900 Quick Start Manual

Corded illumination system
Hide thumbs Also See for KleenSpec 78900:

Advertisement

Cleaning and Disinfecting

CaviWipe
(Wipe)
Cleaning
Enzol (Soak)
Metrizyme
(Soak)
Disinfecting
Caviwipe
* Do not immerse the power supply plug. Follow instructions per this
manual.
Cleaning
Wipe
KleenSpec Corded Illumination System
Refer to Cleaning Table to identify the appropriate cleaning product.
1.
Disconnect the power supply from the wall outlet.
CAUTION: Lamp must remain installed during all cleaning and
disinfecting processes.
2.
Follow the disinfecting wipe manufacturer's instruction and wipe cord or
illuminator completely to remove any residue.
3.
This process is a rough clean only. Refer to disinfection for further
processing.
Soak
Refer to Cleaning Table to identify the appropriate cleaning product.
KleenSpec Corded Illuminator
1.
Disconnect the power supply from the wall outlet.
CAUTION: Lamp must remain installed during all cleaning and
disinfecting processes.
2.
Follow the cleaning solution manufacturer's instructions and soak the
illuminator as instructed. Solution temperature is 22° C to 43° C. Make sure
to keep the power supply out of the soak solution. Do not immerse the
power supply.
3.
Scrub any area of visible debris with a soft brush and thoroughly rinse the
illuminator with 22° C to 43° C tap water to remove the cleaning solution. Do
not rinse the power supply; wipe only.
KleenSpec Corded
Illuminator
Yes
Yes
Yes
Intermediate Level
5
Power supply
Yes
No*
No*
Intermediate
Level

Advertisement

loading

This manual is also suitable for:

Kleenspec 78950