Parallel Connection Of Several Transducers - HBM C2 Series Mounting Instructions

Load cells
Hide thumbs Also See for C2 Series:
Table of Contents

Advertisement

Available languages

Available languages

Connection
Wiring code (six‐wire circuit)
10
(grey)
(black)
(white)
(blue)
(green)
(red)
(-)
Fig. 5.1
Transducer with six lead connection cable
5.3
Parallel connection of several
transducers
Load cells can be wired in parallel by joining the load cell
cable core ends of the same color. In this case, HBM
provides junction boxes of the VKK... series, and for use
in potentially explosive areas, VKK2R‐8 Ex junction
boxes. The output signal is then the average of the indi­
vidual output signals.
Important
Overloading of an individual load cell cannot then be de­
tected from the output signal.
A1572-3.0 HBM: public
Sense (-)
Excitation (-)
Signal (+)
Excitation (+)
Sense (+)
Signal (-)
Shield / wire strand connected with housing
C2A/..., C2A/.../Exd