Automatic Calibration To Room Air; Factors Influencing Calibration - Maxtec MaxO2 OM-25A Operating Manual & Instructions For Use

Table of Contents

Advertisement

continues to occur, contact the service department of the
distributor from which the unit was purchased, or you may call
Maxtec's Customer Service Department directly.

2.2.3 Automatic Calibration to Room Air

The MAXO
analyzer can quickly be calibrated to room air (20.9%)
®
2
using a quick-key shortcut command. This function saves time by
setting the calibration value to 20.9% without scrolling the display. To
use this function:
1. Introduce room air to the sensor at a rate of 1-5 liters per minute
(2 liters per minute is recommended) and allow the reading to
stabilize.
2. Press the LOCK/UNLOCK key
3. Press and hold down the CALIBRATE key
sign starts to flash, press the
value to 20.9%.
4. Release both the CALIBRATE key
The unit will automatically enter the "LOCKED" condition and return
to normal operation.

2.2.4 Factors Influencing Calibration

The primary factors influencing the MAXO
temperature, pressure, and humidity.
Effects of Temperature
The MAXO
analyzer will hold calibration and read correctly within
®
2
+/- 3% when in thermal equilibrium within the operating temperature
range. The device must be thermally stable when calibrated and
allowed to thermally stabilize after experiencing temperature
changes before readings are accurate. For these reasons, the
following is recommended:
1. Allow adequate time for the sensor to equilibrate to a new ambient
temperature.
2. For best results, perform the calibration procedure at a
temperature close to the temperature where analysis will occur.
Pressure Effect
Readings from the MAXO
pressure of oxygen. The partial pressure of Oxygen (PO
the percentage of oxygen (%O
which the sample enviroment is measured. (PO
the readings are proportional to the concentration if the pressure is
held constant. Flow rate of sample gas can affect pressure at the
866.4.Maxtec
arrow key to set the calibration
analyzer are proportional to the partial
®
2
) times the absolute pressure (AP) at
2
www.maxtecinc.com
to unlock the keypad.
. When the "%"
and the
key.
analyzer are
®
2
) is equal to
2
=%O
x AP). Thus
2
2
5