Abbott m-PIMA User Manual page 63

Table of Contents

Advertisement

Error Code
21111-21115, 21120
21501
21502
21503
23020
23028
m-PIMA™ ANALYSER
USER GUIDE
Action
Make sure that the sample capillary is
completely filled. Keep analyser door closed
during test run.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.
Repeat test with new cartridge.
Make sure that the sample capillary is
completely filled.
Too low atmospheric pressure. Do not continue
with testing until you have reached altitudes that
are within specifications.
Too high atmospheric pressure. Do not continue
with testing until you have reached altitudes that
are within specifications.
Cartridge or sample related errors.
Please refer to the test cartridge guides PI-
m-PIMA-04 or PI-m-PIMA-05 for further
information about sample storage and
handling.
Unexpected amplification curve, e.g. due to
individual sample properties.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.