Download Print this page

Emerson Rosemount 5900S Quick Start Manual page 16

Radar level gauge with parabolic antenna
Hide thumbs Also See for Rosemount 5900S:

Advertisement

Quick Start Guide
Figure 4-2: Flange Requirements
A. 116±2 mm
B. 6-38 mm
In case the flange requirements in chapter
are not met, the hole needs to be machined to an oval shape prepared for
inclined welding of the Flange Ball.
Procedure
1. Let the protection plates remain on the Flange Ball until welding is
finished. These plates protect the surface of the Flange Ball from
welding sparks.
16
A. Protection plate
B. Flange Ball
Parabolic antenna requirements
Rosemount 5900S Radar Level Gauge
March 2020

Hide quick links:

Advertisement