Download Print this page

Emerson Rosemount 5900S Quick Start Manual page 31

Radar level gauge with parabolic antenna
Hide thumbs Also See for Rosemount 5900S:

Advertisement

March 2020
5.2
Terminal blocks
Figure 5-2: Rosemount 5900S Terminal Compartment
A
C
+
-
A. Test terminals
B. Ground terminals, internal
C. Field bus
Table 5-1: Terminal Block Connections for the Rosemount 5900S
Connection
X1: Tankbus in
X2: Terminate on
X3: Shield loop
through
X4: Tankbus out
Test terminals
Quick Start Guide
Description
Intrinsically safe Tankbus input, power and communication
(spur in F
Fieldbus system)
OUNDATION
The integrated line terminator is connected over the Tankbus
when a jumper is placed in the terminal block
Cable shield daisy-chain connector (not grounded)
Tankbus output connected to X1 for optional daisy-chain
connection to other devices
Test terminals for temporary connection of a Field
Communicator
Quick Start Guide
A
C
-
+
B
31