ELITE OXYGEN CONSERVER
MODEL: ELH BS PROBE
INSTRUCTION FOR USE
GCE HEALTHCARE
GCE HEALTHCARE
EN

Advertisement

Table of Contents
loading

Summary of Contents for GCE ELITE ELH BS PROBE

  • Page 1 GCE HEALTHCARE GCE HEALTHCARE ELITE OXYGEN CONSERVER MODEL: ELH BS PROBE INSTRUCTION FOR USE...
  • Page 2: Intended Use

    ENGLISH INSTRUCTION FOR USE: ELITE ELH BS PROBE 1. FOREWORD GCE Elite ELH BS PROBE is a medical device classified as class II b according to the Medical Device Directive 93/42/EEC. It´s compliance with the essential requirements 93/42/EEC Medical Device Directive is based upon EN ISO 10524-1 standard.
  • Page 3 If you are in doubt or require any guidance in the use of this product, please contact your distributor or GCE. Before initial use the product should be kept in its original packaging. GCE recommends use of the original packaging (including internal sealing bag and caps) if the product is withdraw from operation (for transport, storage).
  • Page 4: Principle Of Operation

    5. PRINCIPLE OF OPERATION Gas supply: 3.9 bar to 6.0 bar Therapy outlet: Universal ‘fir-tree’ type Therapy flow accuracy: +/- 10% Outlet pressure: 4.2 bar (61 PSI) nominal Oxygen from the cylinder is reduced to a constant pressure and used to fill a reservoir at a rate set by the flow setting knob.
  • Page 5: Use Of Product

    IFU. • Check if the service is due or that the total life time of GCE product has not been exceeded – see Chapter 9. If service or total life time has been exceeded, remove the GCE product from service &...
  • Page 6: Accessories And Spare Parts

    Always use the GCE original spare parts! Accessories: For information about current Accessories, please contact GCE. Spare parts: For information about current spare part list, please contact GCE. 8. CLEANING DO NOT dismantle the unit. DO NOT allow moisture to enter the unit.
  • Page 7 To apply the cleaning solution do not spray it as the spray may enter into the inner parts of product and cause contamination or damage. DO NOT use pressure wash as it could damage or contaminate the product. If the inner parts of the product have been contaminated DO NOT continue to use the product under any circumstances.
  • Page 8 0.1% are included in any GCE product. 9.2. REPAIR AND SERVICE OPERATIONS Repairs and service can be only done by a GCE certified person who also holds all necessary certificates in compliance with national standards for mounting and repair of dedicated gas devices. For information about a service in your area please contact GCE or distributor of GCE product.
  • Page 9: Warranty

    Outlet flow waste. 11. WARRANTY The Standard Warranty period is two years from date of receipt by the GCE Customer (or if this is not known 2 years from time of the product manufac- ture shown on the product). The standard warranty is only valid for products handled according to In-...
  • Page 10 MANUFACTURER: GCE, s.r.o. Tel: +420 569 661 111 Zizkova 381 Fax: +420 569 661 602 583 01 Chotebor http://www.gcegroup.com Czech Republic © GCE, s.r.o.
  • Page 12 IRELAND GERMANY CZECH REPUBLIC FRANCE HUNGARY ROMANIA UNITED STATES ITALY SPAIN OF AMERICA PORTUGAL CHINA MEXICO INDIA PANAMA Manufacturer: GCE, s.r.o. Žižkova 381, 583 01 Chotěboř, Česká republika http://www.gcegroup.com Doc. Nr.: IFU0148 DOI: 2020-08-26; Rev.: 00; TI: A6, CB, V1...