Meter Configuration
The digital reader for the Field Scout EC meter must
be in Manual, 1-Point calibration mode to allow the
meter to be calibrated with the 2.76 mS/cm calibrat-
ing solution. The meter should also be set to the HI
range mode. This is how the meters are shipped.
Configuring the calibration mode
1. Switch off the display.
2. Press and hold the INC key (see p. 6 for diagram
of calibration keys) then simultaneously press the
ON/OFF button to switch on the display. The meter
will go into calibration selection mode.
3. The lower display shows "A.CAL" and the upper
display blinks the current choice (YES or NO). Press
either the INC or DEC key to select the NO option.
This disables the Automatic calibration mode and
configures the display for Manual calibration.
4. Press the HOLD/ENTER button to confirm the se-
lection. The display shows "CO".
5. After 1 second, the lower display will show "1.Pnt"
and the upper display blinks the current choice (YES
or NO). Press either the INC or DEC key to select
the YES option. This enables the 1-Point calibration
option.
4. Press the HOLD/ENTER button to confirm the se-
lection. The display shows "CO" and, after a few
seconds, goes to measurement mode.