Operation: As a Body Thermometer
1. Press the ON/OFF button.
2. When the preparation screen displays (see above) the device is ready for
temperature measurement.
3. Align the device with the centre of the forehead, ensuring the distance between the
forehead and the device is between 10mm and 30mm.
4. Press the ON/OFF button to take temperature. The temperature will appear on the
display, accompanied by one long beep.
If the reading is between 37.5°C(99.5°F) and 43°C(109.4°F), the display will be accompanied
by ten short beeps.
As forehead temperature measurement could be affected by sweat, oil and the ambient
temperature of the location, the reading should be taken as a reference only.
If the probe is placed at an angle close to the forehead, the reading will be affected by
surrounding temperature.
Babies' skin reacts very quickly to ambient temperature. Therefore, do not take their
temperature during/after breastfeeding as the skin temperature may be lower than their
internal body temperature.
The user and thermometer must be in the room where temperature is to be taken for 30
minutes prior to temperature being taken.
Operation: As an Object Thermometer
1. Press the ON/OFF button.
2. Ensure the device is in Object Mode (see Page 7)
3. Point the device at the object to be measured.
4. Press the ON/OFF button to take temperature. The temperature will appear on the
display, accompanied by one beep.
As forehead temperature measurement could be affected by sweat, oil and the ambient
temperature of the location, the reading should be taken as a reference only.
If the probe is placed at an angle close to the object, the reading will be affected by
surrounding temperature.