marsden UFR103 User Manual

marsden UFR103 User Manual

Non contact infrared thermometer
Hide thumbs Also See for UFR103:

Advertisement

UFR103 Non Contact
Infrared Thermometer
User Manual
-
Version 1.2 09/20

Advertisement

Table of Contents
loading

Summary of Contents for marsden UFR103

  • Page 1 UFR103 Non Contact Infrared Thermometer User Manual Version 1.2 09/20...
  • Page 2: Table Of Contents

    Contents Introduction Declaration of Conformity Specification Symbol Descriptions Important Safety Instructions Parts Identification Battery Installation Mode Setting Unit & Sound Setting About Normal Temperature & Fever Measurement Ranges Operation Error Messages Warranty EMC Declaration...
  • Page 3: Introduction

    Thank you for purchasing this Marsden UFR103 Non Contact Infrared Thermometer. This thermometer uses advanced infrared (IR) technology to measure forehead or object temperature instantly and accurately. To ensure accurate use of the UFR103, please read this user manual before use and keep to hand for future reference.
  • Page 4: Specification

    Specification Measurement Range/Accuracy Body mode:32.0℃~43.0℃(89.6℉~109.4℉) Object mode:0.0℃~100.0℃(32.0℉~199.9℉) Ambient temperature:0.0℃~40.0℃(32.0℉~104.0℉) 10mm – 30mm Measuring Distance ℃/℉ Temperature Unit Display Resolution 0.1℃/0.1℉ Accuracy ±0.2°C/±0.4°F (within 35.0°C~42.0C/ 95.0°F~107.6°F) Memory Function 20 sets of measurement values Buzzer Function (1)Turn on the device:1Short beep (2) Measurement completed: 1 long beep (3) Fever>...
  • Page 5: Important Safety Instructions

    Important Safety Instructions Before using this device, please read the following instructions with care. WARNING: • This thermometer is not intended to substitute for a consultation with your physician. The forehead scan temperature serves as a reference only. • Basic safety precautions should always be observed, especially when the thermometer is used on or near children and disabled persons.
  • Page 6: Parts Identification

    Parts Identification Battery Installation 1. Remove the battery cover from the battery compartment. 2. Insert two 1.5V AAA batteries ensuring each one is facing the correct way. Positive (+) and Negative (-) are displayed on the back of battery cover. 3.
  • Page 7: Mode Setting

    Mode Setting 1. Switch the scale on using the ON/OFF button. 2. Default start-up setting for the device is Body Mode. Short press the SET button once to switch to Object Mode. 3. Press the SET button again to switch to Ambient Temperature Mode. 4.
  • Page 8: About Normal Temperature & Fever

    Before Use: About Normal Body Temperature & Fever Forehead and temple area temperature differs from internal temperature, which can be taken orally or rectally. Vasoconstriction, an effect which constricts the blood vessels and cools the skin, can occur during the early stages of a fever.
  • Page 9: Operation

    Operation: As a Body Thermometer 1. Press the ON/OFF button. 2. When the preparation screen displays (see above) the device is ready for temperature measurement. 3. Align the device with the centre of the forehead, ensuring the distance between the forehead and the device is between 10mm and 30mm.
  • Page 10 Operation: As an Ambient Thermometer 1. Press the ON/OFF button. 2. Ensure the device is in Ambient Temperature Mode (see Page 7) 3. Press the ON/OFF button to take temperature. The temperature will appear on the display, accompanied by one beep. Memory Recall of Measurements 1.
  • Page 11: Error Messages

    Error Messages Symbol Explanation In Body Mode, measured temperature is above the measuring range of 43 ℃/109.4 ℉. In Body Mode, measured temperature is below the measuring range of 32℃/89.6 ℉ In Object Mode, measured temperature is above the measuring range 100.0℃/199.9 or environmental temperature is above ℉,...
  • Page 12: Warranty

    Warranty • The device is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the date of purchase. • For repair under this warranty, our authorised service agent must be advised of the fault within the period of the warranty.
  • Page 13: Emc Declaration

    EMC Declaration IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product Instructions for use The ME EQUIPMENT or ME SYSTEM is suitable for home and healthcare environments. Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
  • Page 14 Table 2 Guidance and manufacturer’s declaration - electromagnetic Immunity Immunity Test IEC 60601-1-2 Test level Compliance level Electrostatic discharge (ESD) IEC ±8 kV contact ±2 kV, ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV 61000-4-2 ±4 kV, ±8 kV, ±15 kV Electrical fast transient/burst IEC Power supply lines: ±2 kV input/output lines: 61000-4-4...
  • Page 15 Table 3 Guidance and manufacturer’s declaration - electromagnetic Immunity Test IMMUNITY Band Modulation Distance Frequency Service Modulation TEST (MHz) (MHz) LEVEL(V/m) Pulse 380 – TETRA 400 modulation 18 Hz FM ± 5 kHz 380 – GMRS 460, FRS deviation 1 kHz sine Radiated RF Pulse...
  • Page 16 +44 (0) 1709 364296 sales@marsdengroup.co.uk Distribution: Unit 7, Centurion Business Park, Coggin Mill Way, Rotherham, S60 1FB Head Office: Unit 1, Genesis Business Park, Sheffield Road, Rotherham S60 1DX www.marsden-weighing.co.uk...