Emerson Fisher LCP100 Instruction Manual page 6

Local control panel
Hide thumbs Also See for Fisher LCP100:
Table of Contents

Advertisement

LCP100 Local Control Panel
August 2012
Figure 3. Wiring for 24 VDC External Power Configuration
LOGIC
SOLVER
OUTPUT
4‐20 mA
(USER SUPPLIED)
24 VDC
SOURCE
(USER SUPPLIED)
NOTE: DO NOT CONNECT THE LOOP + TERMINAL IN THE LCP100 TO THE LOOP + TERMINAL IN THE DVC6200 SIS.
THIS WILL CAUSE THE LCP100 TO UNNECESSARILY CONSUME 4 mA AT THE EXPENSE OF THE DVC6200 SIS.
1 THIS CONNECTION IS ALSO LABELED LOOP -.
E1465
6
LCP100
SWITCH TO
24VDC POSITION
24VDC +
24VDC *
SIMPLE METHOD FOR INSTALLING AN LCP100 TO AN EXISTING
DVC6200 SIS INSTRUMENT WHEN 24 VDC POWER IS AVAILABLE
AUX +
CASE GROUND
1
AUX*
SHIELD NOT CONNECTED TO LCP100
Instruction Manual
D103272X012
24VDC
DVC6200 SIS TERMINAL BOX