13. Non-Contact Voltage (NCV) Detection (Picture 16)
Picture 16
1) Turn the function dial to the
NCV
position.
2) Bring the NCV detector (top left corner of the meter) close to the wire (AC)
under test.
3) If the voltage of the wire is ≥50Vrms (frequency: 50Hz/60Hz), the red indicator
light will be on and the buzzer will beep. If no voltage is detected, the LCD will
display "EF". As the intensity of the detected voltage increases, more segments
"-" will be displayed, and the frequency for buzzer beeping and red indicator
light
flashing will be higher.
Caution:
The detected voltage level varies with the distance between the NCV detector
and the
wire
under test.
The detected voltage level is for reference only, not for specific measurement.
The frequency of the detected voltage should be 50Hz/60Hz.
Hold the meter casing for non-contact voltage detection.
23
FUSED
250V max
MAX 10sec
EACH 15min
14. USB Data Transmission (Picture 17a, Picture 17b)
Picture 17a
1) Pull out the USB sealing cover at the back of the meter (Picture 17a).
2) Insert the USB communication module into the USB access port of the meter
and the LCD will display "
" (Picture 17b).
3) If USB data transmission is not needed during measurement, long press the
button or pull out the USB module to disable data transmission, and
"
" will disappear.
4) To recover this function, long press the
5) The USB communication software can be downloaded from the official website
of Uni-Trend (www.uni-trend.com).
15. Others
1) Auto power off: During measurement, if there is no operation for 15 minutes,
the meter will automatically shut down to save power; before the automatic
shutdown, the buzzer will beep for warning. Users can wake the meter up by
pressing the
SELECT
button. To disable the auto-off function, press and hold
the
SELECT
button in the off state, and then turn on the meter. To recover the
function, restart the meter.
2) Buzzer alarm during measurement: When the input voltage >1000V or current
>10A, the buzzer will sound an alarm.
3) Low battery indication: When the battery voltage is ≤4.6V ±0.2V, "
be displayed.
Picture 17b
button or insert the USB module.
" will
24