ELECTRICAL CONNECTIONS
3
3.1.2 Remote version
Terminals for electrical installation
Figure 3-5: Terminals for electrical installation
1 Grounding terminal in the housing (if the electrical cable is shielded)
2 Current output -
3 Current output +
4 Location of the external grounding terminal (on the wall support)
INFORMATION!
Electrical power to the output terminal energizes the device. The output terminal is also used for
HART
communication.
®
CAUTION!
Use the applicable electrical cables with the cable glands.
•
Make sure that the current is not more than 5 A or that there is 5 A-rated fuse in the
•
electrical circuit that energizes the device.
Connections between the remote converter and the probe housing
Figure 3-6: Connections between the remote converter and the probe housing
1 Remote converter
2 Probe housing
3 Power supply: voltage in -
4 Power supply: voltage in +
5 Signal cable B
6 Signal cable A
7 Shielding wire (attached to Faston connectors in the housings of the remote converter and the probe housing)
For more electrical installation data, refer to
30
Compact version
www.krohne.com
POWERFLEX 2200 C/F/S/D
on page 28.
07/2016 - 4005318901 - QS POWERFLEX 2200 R01 en