LaserPoint Plus 2 Quick Start Manual page 4

Power & energy meter
Hide thumbs Also See for Plus 2:

Advertisement

2.5. M
ANUAL OR AUTOMATIC POWER RANGING SELECTION
The measurement full scale or range can be adjusted according to user's needs. The selection can be done
by touching the "Range" screen button
Select the appropriate range manually, or choose the "Auto" mode
meter managing the measurement range automatically.
2.6. O
FFSET FUNCTION
In certain cases it is useful to introduce an offset level in a measurement. To activate this function
touch the "Zero" screen button
acquired as offset level. When the "Offset" function is activated, the offset value/units appear in
the "Zero" screen button and an additional yellow spot is shown at the upper right corner of the
"Zero" screen button
2.7. Z
S
ERO
ETTINGS
All meter adjustments, including ADC zeroing, are carried out by Plus 2 firmware. However
Laserpoint recommend to reset Plus 2 Meter every time a new sensor head is being plugged in
following the instructions given below:
a)
Press for more the Zero button
the window shown in Figure 5 appears in the
display segment e) as shown in Figure
Note:
While zeroing make sure the instrument is not affected
by electrical noise and the sensor is blind to light.
b) Press "Yes" and wait few seconds whilel the display
shows the notice "Wait for zeroing". When the ADC
has been reset the notice "Zeroing Completed" is
finally displayed (Figure 6) and the meter is ready
again to be used.
Laserpoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved
:
. To cancel the offset, press the "Zero" button again.
1.
:
and scrolling to choose the range option.
and the presently measured power value is
until
to let Plus 2
Figure 5
Figure 6
4