Download Print this page

LaserPoint Plus 2 Quick Start Manual

Power & energy meter
Hide thumbs Also See for Plus 2:

Advertisement

Quick Links

Laserpoint srl - Via Burona, 51 - 20090 Vimodrone (Milano) - Italy
Phone +39 02 27 400 236 - Fax +39 02 25 029 161-www.laserpoint.eu
1
LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved

Advertisement

loading

Summary of Contents for LaserPoint Plus 2

  • Page 1 Laserpoint srl - Via Burona, 51 - 20090 Vimodrone (Milano) - Italy Phone +39 02 27 400 236 - Fax +39 02 25 029 161-www.laserpoint.eu LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved...
  • Page 2: Quick Start

    TART 2.1. T ISPLAY As shown in Figure 1, the touch screen display of Plus 2 is divided in horizontal sections with a specific content: a) Date - Time - Battery/network icons b) Sensor Head name and Serial number; Sensor Head temperature (°C)
  • Page 3 Plus 2 will turn off in a few seconds (while in battery mode). To turn off the Plus 2 press the on-off button for 2-3 seconds, then release it. The current Plus 2 configuration will be automatically saved.
  • Page 4 2.7. Z DJUSTMENTS In the Plus 2, all adjustments, including ADC zeroing, are done by the software. It is recommended to re-zero the Plus 2 each time the measuring head is changed. Press for more than 3 seconds the Zero button on the screen of the Plus 2.
  • Page 5 EASURING ENERGY OF A SINGLE PULSE OR BURST PULSE Set the Energy Mode as described above. The green led near the measured value means that the Plus 2 is ready to measure the energy pulse. When firing the laser, the value display will turn blank, the LED will turn red and the writing “Acquisition”...
  • Page 6 ), while in Energy mode they’ll be displayed as Fluence (unit of measure: J/cm Figure 12 Note: to deactivate Irradiance/Fluence measures Open the “Measure Setting” menu, and select the Shape “none” to deactivate. LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved...
  • Page 7 Manual acquisition: check this box to start and stop the acquisition manually Timed acquisition: uncheck “Manual acquisition” box and input the duration after which the acquisition will be stopped. LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved...
  • Page 8: The "Menu " Button

    ON/OFF activate/deactivate, and set a time from 1 to 30 minutes. d) Auto off is the inactivity time before the Plus 2 will turn off. You can select ON/OFF to activate/deactivate, and set a time Figure 18 up to 600 minutes.
  • Page 9 90° FullScreen: switch to a full screen high visibility / high contrast display (Figure 19) showing only the measure digits and the unit of measure. Click the “X” button to return to main screen. Figure 19 LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved...
  • Page 10 > 15 hours with stand-by display option Input 100/240Vac 50/60Hz Out 5Vdc 1A, Charging current is 0.5A (Plus 2 may be charged Supplied Charger through PC USB port). LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved...