Anritsu VectorStar ME7848A Series Maintenance Manual

Anritsu VectorStar ME7848A Series Maintenance Manual

Opto-electronic network analyzer
Hide thumbs Also See for VectorStar ME7848A Series:
Table of Contents

Advertisement

Quick Links

Maintenance Manual
TM
VectorStar
ME7848A Series
Opto-electronic Network Analyzer
ME7848A-0240 40 GHz, 850 nm system
ME7848A-0270 70 GHz, 1550 nm system
ME7848A-0271 70 GHz, 1310 nm system
ME7848A-0140 40 GHz, 850 nm system (VNA and O/E module only)
ME7848A-0170 70 GHz, 1550 nm system (VNA and O/E module only)
ME7848A-0171 70 GHz, 1310 nm system (VNA and O/E module only)
Anritsu Company
P/N: 10410-00778
490 Jarvis Drive
Revision A
Morgan Hill, CA 95037-2809
Published: December 2019
USA
Copyright 2019 Anritsu Company. All Rights Reserved.
http://www.anritsu.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VectorStar ME7848A Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Anritsu VectorStar ME7848A Series

  • Page 1 ME7848A-0170 70 GHz, 1550 nm system (VNA and O/E module only) ME7848A-0171 70 GHz, 1310 nm system (VNA and O/E module only) Anritsu Company P/N: 10410-00778 490 Jarvis Drive Revision A Morgan Hill, CA 95037-2809 Published: December 2019 Copyright 2019 Anritsu Company. All Rights Reserved. http://www.anritsu.com...
  • Page 3: Table Of Contents

    VectorStar ME7838 Series 2-Port BB/mmW VNA Measurement System....1-2 VectorStar ME7848A Series Opto-electronic Network Analyzer ..... 1-2 VectorStar™...
  • Page 4 ME7848A-0270 1550 nm Test Records ......... . A-5 Contents-2 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 5: Chapter 1 - General Information

    From here, you can select the latest sales, service and support contact information in your country or region, provide online feedback, complete a “Talk to Anritsu” form to get your questions answered, or obtain other services offered by Anritsu.
  • Page 6: Related Documentation

    VectorStar MN4765B O/E Calibration Module • MN4765B O/E Calibration Module Operating Manual – 10410-00742 Updates to Manuals For updates to any of the VectorStar Series VNA documentation, visit Anritsu’s Web site at: http://www.anritsu.com/en-us/test-measurement/products/ms4640b-series PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 7: Electrostatic Discharge (Esd) Prevention

    • Includes front panel loops above with four (4) attenuators, bias tees in test set, gain compression, and efficiency measurement software. 2000-1957-R 850 nm Accessory Kit for 40 GHz ME7848A ONA system VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 8 Includes front panel loops above with four (4) attenuators, bias tees in test set, gain compression, and efficiency measurement software. 2000-1958-R 1310 and 1550 nm Accessory Kit for 70 GHz ME7848A ONA system PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 9: Me7848A-01Xx Standard Broadband Vna System Components

    Includes front panel loops above with four (4) attenuators, bias tees in test set, gain compression, and efficiency measurement software. 2000-1958-R 1310 and 1550 nm Accessory Kit for 70 GHz ME7848A ONA system VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 10: Recommended Test Equipment

    Critical Specification Manufacturer and Model Codes Optical Power Meter ThorLabs PM100D P, T Photodiode Power Sensor Thorlabs S155C P, T a.Use Codes: P = Performance Verification; A = Adjustment; T = Troubleshooting PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 11: Chapter 2 - Replaceable Parts

    There are no serviceable components or subassemblies inside the MN4775A-0040, Note MN4775A-0070, and MN4775A-0071 E/O converters or the MN4765B-0040, MN4765B-0070, and MN4765B-0071 O/E modules. The modules must be returned to Anritsu Company for repair. Table 2-1. ME7848A Series System Level Replaceable Parts List...
  • Page 12 2-2 ME7848A Series System Replaceable Parts Replaceable Parts PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 13: Chapter 3 - Performance Verification

    If equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Only with written consent from Anritsu may changes to single components be carried out or components not supplied by Anritsu be used.
  • Page 14: Calibration And Measurement Conditions

     both during calibration and during measurement. • Warm-up Time: • 90 minutes • Environmental Conditions • Temperature • 23 °C ± 3 °C, with < 1 °C variation from calibration temperature • Relative Humidity • 20-50% recommended PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 15: Me7848A-0240 System Verification - 850 Nm Configuration

    ME7848A-0240 850 nm System Setup 1. Configure components (VNA, O/E, E/O, RF, and Optical cables) as shown in Figure 3-1. The Anritsu MN4775A-0040 does not have Laser Out and Laser In connectors like the Note MN4775A-0070 and MN4775A-0071. Anritsu MN4775A-0040 Ignition Key “OFF”...
  • Page 16 Touch VOA ON square-pad to enable VOA Attenuator. c. Touch BIAS ON square-pad to enable BIAS Regulator. 10. Select ITU-Channel. • Not required for 850 System as wavelength is fixed at 852 nm. PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 17 At Secondary Screen select MODE = QUADRATURE. 13. Typical default BIAS SETTINGS MN4775A-0040: a. Dither-Frequency 3 kHz b. Dither-Amplitude 600 mVpp. c. Dither ON. d. Dither Slope POSITIVE. 14. Turn E/O Converter LASER Inter-Lock Ignition Key to OFF. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 18: Optical Power Meter (Opm) Measurement Setup - 850 Nm Configuration

    On Front Panel select Lambda button as shown in Figure 3-3. b. Navigate to 850 nm Wavelength icon in menu bar area. c. On PM100D Meter select OK to choose 850 nm selection, as shown in Figure 3-3. PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 19: Optical Power Meter Measurement - 850 Nm Configuration

    Meter reading may fluctuate nominally ± 0.5 dB. Note 2. E/O Converter Inter-Lock to OFF: a. Set E/O Converter Gen Laser Interlock to OFF via Ignition Key. b. Power down Optical Power Meter. 3. Disconnect the Optical Power Meter. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 20: Frequency Response Repeatability S21 Setup - 850 Nm Configuration

    4. Connect Green-colored end of Optical cable to MN4765B-0040 as shown in Figure 3-5. 5. Connect the VNA Port 1 to Anritsu MN4775A-0040 E/O Converter RF Input as shown in Figure 3-5. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 21 Dither Slope POSITIVE SLOPE. e. Touch Home to return to the main screen. 13. Preset the VectorStar VNA as follows: a. Select Preset button on the icon bar and then the OK button. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 22: Frequency Response Repeatability S21-Mag Measurement - 850 Nm Configuration

    7. Copy the twenty five (25) saved data files off the MS464xB onto a USB flash drive for transferring to a Personal Computer. 8. On a separate Windows Personal Computer, import the saved data from the HLN#n files into Microsoft Excel so the RMS values can be calculated. 3-10 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 23 Record the largest calculated RMS value of each frequency band listed in the Frequency Response S21 worksheet into Table A-2, “Frequency Response Repeatability – 850 nm Band” on page A-1. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-11...
  • Page 24: Optical Noise Floor S21 Setup - 850 Nm Configuration

    4. Connect Green-colored end of optical cable to MN4765B-0040 as shown in Figure 3-7. 5. Connect the VNA Port 1 to Anritsu MN4775A-0040 E/O Converter RF Input as shown in Figure 3-7. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 25 In this new screen, enter the information from the first entry in Figure 3-8. d. Select Add and repeat until all the segments are entered. e. Set Sweep type to Segmented Sweep (Freq Based). Figure 3-8. MS4644B Frequency -Based Segmented Sweep Setting VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-13...
  • Page 26: Optical Nose Floor S21 Measurement - 850 Nm Configuration

    3. Select Manual Cal. 4. Select Transmission Freq. Response. 5. Select Modify Cal Setup. 6. Select Edit Cal Params (see Figure 3-9). Figure 3-9. TRANSMISSION FREQUENCY RESPONSE CAL SETUP (SOLT/R, Coaxial) Dialog 3-14 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 27 Calibration lamp in VNA is ON: a. Select Thru/Recip. b. Select Thru and wait for green checkmark to appear. c. Select Back. d. Select Done. e. Verify Green Calibration lamp is ON. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-15...
  • Page 28 16. Copy the sixteen (16) saved data files off the MS464xB onto a USB flash drive for transferring to a Personal Computer. 17. On a separate Windows Personal Computer, import the saved data from the NF#n files into Microsoft Excel so the RMS values can be calculated. 3-16 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 29 Table A-3, “Optical Noise Floor S21 (Option 51) – 850 nm Band” on page A-2 Table A-4, “Optical Noise Floor S21 (Option 61 or 62) – 850 nm Band” on page A-2. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-17...
  • Page 30: Me7848A-0271 System Verification - 1310 Nm Configuration

    Anritsu MN4765B-0071 Figure 3-12. Initial ME7848A-0271 1310 nm Initial System Setup Refer to the MN4775A Operation Manual – 10410-00774 regarding proper operation of the Anritsu Note E/O Converter. 2. Apply power to MN4775A-0071 E/O Converter by pushing the Standby button until it turns green.
  • Page 31 Touch VOA ON square-pad to enable VOA Attenuator. c. Touch BIAS ON square-pad to enable BIAS Regulator. 10. Select ITU-Channel. • Not required for 1310 System as wavelength is fixed. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-19...
  • Page 32 13. Typical default BIAS SETTINGS MN4775A-0071: a. Dither-Frequency 3 kHz. b. Dither-Amplitude 600 mVpp. c. Dither ON. d. Dither Slope POSITIVE. 14. Turn E/O Converter LASER Inter-Lock Ignition Key to OFF. 3-20 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 33: Optical Power Meter (Opm) Measurement Setup - 1310 Nm Configuration

    On Front Panel select Lambda button as shown in Figure 3-14. b. Navigate to 1310 nm Wavelength icon in menu bar area. c. On PM100D Meter select OK to choose 1310 nm selection, as shown in Figure 3-14. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-21...
  • Page 34: Optical Power Meter Measurement - 1310 Nm Configuration

    Note 2. E/O Converter Inter-Lock to OFF: a. Set E/O Converter Gen Laser Inter-Lock to OFF via Ignition Key b. Power down Optical Power Meter 3. Disconnect the Optical Power Meter. 3-22 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 35: Frequency Response Repeatability S21 Setup - 1310 Nm Configuration

    4. Connect Green-colored end of Optical cable to MN4765B-0071 as shown in Figure 3-16. 5. Connect the VNA Port 1 to Anritsu MN4775A-0071 E/O Converter RF Input as shown in Figure 3-16. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 36 Dither Slope POSITIVE SLOPE. e. Touch Home to return to the main screen. 14. Preset the VectorStar VNA as follows: a. Select Preset button on the icon bar and then the OK button. 3-24 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 37: Frequency Response Repeatability S21-Mag Measurement - 1310 Nm Configuration

    7. Copy the twenty five (25) saved data files off the MS465xB onto a USB flash drive for transferring to a Personal Computer. 8. On a separate Windows Personal Computer, import the saved data from the HLN#n files into Microsoft Excel so the RMS values can be calculated. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-25...
  • Page 38 Record the largest calculated RMS value of each frequency band listed in the Frequency Response S21 worksheet into Table A-6, “Frequency Response Repeatability – 1310 nm Band” on page A-3. 3-26 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 39: Optical Noise Floor S21 Setup - 1310 Nm Configuration

    4. Connect Green-colored end of Optical cable to MN4765B-0071 as shown in Figure 3-18. 5. Connect the VNA Port 1 to Anritsu MN4775A-0071 E/O Converter RF Input as shown in Figure 3-18. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 40 In this new screen, enter the information from the first entry in Figure 3-19. d. Select Add and repeat until all the segments are entered. e. Set Sweep type to Segmented Sweep (Freq Based). Figure 3-19. MS4647B Frequency-Based Segmented Sweep Settings 3-28 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 41: Optical Nose Floor S21 Measurement - 1310 Nm Configuration

    3. Select Manual Cal. 4. Select Transmission Freq. Response. 5. Select Modify Cal Setup. 6. Select Edit Cal Params (see Figure 3-20). Figure 3-20. TRANSMISSION FREQUENCY RESPONSE CAL SETUP (SOLT/R, Coaxial) Dialog VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-29...
  • Page 42 Calibration lamp in VNA is ON. a. Select Thru/Recip. b. Select Thru and wait for green checkmark to appear. c. Select Back. d. Select Done. e. Verify Green Calibration lamp is ON. 3-30 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 43 11. Set E/O Converter Ignition key to OFF (see figure Figure 3-22). Anritsu MN4775A-0071 Ignition Key “OFF” Laser Out/In Optical Loop Connected VectorStar Port 1 Port 2 Anritsu MN4765B-0071 Figure 3-22. Turning Off Laser VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-31...
  • Page 44 16. Copy the sixteen (16) saved data files off the MS465xB onto a USB flash drive for transferring to a Personal Computer. 17. On a separate Windows Personal Computer, import the saved data from the NF#n files into Microsoft Excel so the RMS values can be calculated. 3-32 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 45 Table A-7, “Optical Noise Floor S21 (Option 51) – 1310 nm Band” on page A-4 Table A-8, “Optical Noise Floor S21 (Option 61 or 62) – 1310 nm Band” on page A-4. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-33...
  • Page 46: Me7848A-0270 System Verification - 1550 Nm Configuration

    Anritsu MN4765B-0070 Figure 3-24. Initial ME7848A-0270 1550 nm Initial System Setup Refer to the MN4775A Operation Manual – 10410-00774 regarding proper operation of the Anritsu Note E/O Converter. 2. Apply power to MN4775A-0070 E/O Converter by pushing the Standby button until it turns green.
  • Page 47 (left side). a. Touch LASER ON square-pad to turn the LASER to ON state. b. Touch VOA ON square-pad to enable VOA Attenuator. c. Touch BIAS ON square-pad to enable BIAS Regulator. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-35...
  • Page 48 13. Typical default BIAS SETTINGS MN4775A-0070: a. Dither-Frequency 3 kHz. b. Dither-Amplitude 600 mVpp. c. Dither ON. d. Dither Slope POSITIVE. 14. Turn E/O Converter LASER Inter-Lock Ignition Key to OFF. 3-36 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 49: Optical Power Meter (Opm) Measurement Setup - 1550 Nm Configuration

    On Front Panel select Lambda button as shown in Step 3-26. b. Navigate to 1550nm Wavelength icon in menu bar area. c. On PM100D Meter select OK to choose 1550 nm selection, as shown in Figure 3-26. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-37...
  • Page 50: Optical Power Meter Measurement - 1550 Nm Configuration

    2. Set E/O Converter Inter-Lock to OFF: a. Set E/O Converter Gen Laser Interlock to OFF via Ignition Key. b. Power down Optical Power Meter. 3. Disconnect the Optical Power Meter. 3-38 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 51: Frequency Response Repeatability S21 Setup - 1550 Nm Configuration

    4. Connect Green-colored end of Optical cable to MN4765B-0070 as shown in Figure 3-28. 5. Connect the VNA Port 1 to Anritsu MN4775A-0070 E/O Converter RF Input as shown in Figure 3-28. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 52 Dither Slope POSITIVE SLOPE. e. Touch Home to return to the main screen. 14. Preset the VectorStar VNA as follows: a. Select Preset button on the icon bar and then the OK button. 3-40 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 53: Frequency Response Repeatability S21-Mag Measurement - 1550 Nm Configuration

    7. Copy the twenty five (25) saved data files off the MS464xB onto a USB flash drive for transferring to a Personal Computer. 8. On a separate Windows Personal Computer, import the saved data from the HLN#n files into Microsoft Excel so the RMS values can be calculated. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-41...
  • Page 54 Rename and typing in the new name. f. Record the largest calculated RMS value of each frequency band listed in Frequency Response S21 worksheet into Table A-10, “Frequency Response Repeatability – 1550 nm Band” on page A-5. 3-42 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 55: Optical Noise Floor S21 Setup - 1550 Nm Configuration

    4. Connect Green-colored end of Optical cable to MN4765B-0070 as shown in Figure 3-30. 5. Connect the VNA Port 1 to Anritsu MN4775A-0070 E/O Converter RF Inputas shown in Figure 3-30. • Use torque wrench to tighten the connectors to ensure that the connections do not work themselves loose during the test.
  • Page 56 Select Freq-based Seg. Sweep Setup and a new window appears at the bottom of the screen. c. In this new screen, enter the information from the first entry in Figure 3-31. d. Select Add and repeat until all the segments are entered. 3-44 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 57 Performance Verification 3-6 ME7848A-0270 System Verification – 1550 nm Configuration e. Set Sweep type to Segmented Sweep (Freq Based). Figure 3-31. MS4647B Frequency-Based Segmented Sweep Settings VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-45...
  • Page 58: Optical Nose Floor S21 Measurement - 1550 Nm Configuration

    3. Select Manual Cal. 4. Select Transmission Freq. Response. 5. Select Modify Cal Setup. 6. Select Edit Cal Params (see Figure 3-32). Figure 3-32. TRANSMISSION FREQUENCY RESPONSE CAL SETUP (SOLT/R, Coaxial) Dialog 3-46 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 59 Calibration lamp in VNA is on. a. Select Thru/Recip. b. Select Thru and wait for green checkmark to appear. c. Select Back. d. Select Done. e. Verify Green Calibration lamp is ON. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-47...
  • Page 60 11. Set E/O Converter Ignition key to OFF, see figure Figure 3-34. Anritsu MN4775A-0070 Ignition Key “OFF” Laser Out/In Optical Loop Connected VectorStar Port 1 Port 2 Anritsu MN4765B-0070 Figure 3-34. Turning Off Laser 3-48 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 61 16. Copy the sixteen (16) saved data files off the MS464xB onto a USB flash drive for transferring to a Personal Computer. 17. On a separate Windows Personal Computer, import the saved data from the NF#n files into Microsoft Excel so the RMS values can be calculated. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A 3-49...
  • Page 62 Table A-11, “Optical Noise Floor S21 (Option 51) – 1550 nm Band” on page A-6 Table A-12, “Optical Noise Floor S21 (Option 61 or 62) – 1550 nm Band” on page A-6. 3-50 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 63: Chapter 4 - Theory Of Operation

    • MN4775A E/O converter • MN4765B O/E calibration module Figure 4-1 on page 4-2 shows the ME7848A Series ONA system configuration and illustrates the interconnections among the VNA, E/O converter, and O/E calibration module. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 64 4-3 System Components Theory of Operation Figure 4-1. ME7848A Series ONA System Interconnections PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 65: Functional Description Of System Components

    < 2.5 GHz Low Band Bias 1 Bias 2 Test Test Port 2 Port 1 1. Items with broken (dashed) lines are optional connections. Figure 4-2. VectorStar MS4640B Series VNA Simplified Block Diagram VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 66: Mn4775A E/O Converter

    The MN4765B is a characterized, unamplified photodiode module. It is used as an optical receiver with the Anritsu MS464xB and MS4652xB Series VNAs to perform highly accurate and stable optoelectronic measurements of both modulators (E/O) and photoreceivers (O/E). Bandwidth and wavelength coverage depends on the option selected when the module is ordered.
  • Page 67: Me7848A Ona System Operation

    The test signal from the MN4765B RF connector is then fed back to the test port of the MS464xB VNA for further signal processing. In the MS4640B Series Microwave Vector Network Analyzer Maintenance Manual - Note 10410-00320, refer to Chapter 2 - Theory of Operation for signal processing details. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 68 4-5 ME7848A ONA System Operation Theory of Operation PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 69: Chapter 5 - Troubleshooting

    If equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Only with written consent from Anritsu may changes to single components be carried out or components not supplied by Anritsu be used.
  • Page 70 This instrument should be kept clear of environments where liquid spills or condensing moisture are likely. It is not water resistant. To avoid damage to the instrument, do not expose it to spray, liquids, or solvents. PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 71 However, there is no guarantee that interference will not occur in a particular installation. VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 72: Troubleshooting Strategy

    Turn the laser on and off to verify that the level changes. 5. If you have low power, clean all of optical cables and optical connectors. Also, verify the RF coaxial cables are clean and correctly torqued. PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 73: Appendix A - Me7848A Ona Test Records

    MN4775A-0040 0.014 Band (0.300-00002 MHz) MN4775A-0040 0.014 Band (00002-00010 MHz) MN4775A-0040 0.01 Band (00010-02500 MHz) MN4775A-0040 0.01 Band (02500-05000 MHz) MN4775A-0040 0.01 Band (05000-20000 MHz) MN4775A-0040 0.012 Band (20000-40000 MHz) MN4775A-0040 0.02 VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 74 MN4775A-0040 0.42 Band (0.300-00002 MHz) MN4775A-0040 0.49 Band (00002-00010 MHz) MN4775A-0040 0.42 Band (00010-02500 MHz) MN4775A-0040 0.58 Band (02500-05000 MHz) MN4775A-0040 0.46 Band (05000-20000 MHz) MN4775A-0040 0.46 Band (20000-40000 MHz) MN4775A-0040 0.58 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 75: Me7848A-0271 1310 Nm Test Records

    MN4775A-0071 0.01 Band (05000-20000 MHz) MN4775A-0071 0.01 Band (20000-38000 MHz) MN4775A-0071 0.01 Band (38000-50000 MHz) MN4775A-0071 0.01 Band (50000-65000 MHz) MN4775A-0071 0.01 Band (65000-67000 MHz) MN4775A-0071 0.01 Band (67000-70000 MHz) MN4775A-0071 0.02 VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 76 MN4775A-0071 0.46 Band (05000-20000 MHz) MN4775A-0071 0.46 Band (20000-38000 MHz) MN4775A-0071 0.58 Band (38000-50000 MHz) MN4775A-0071 0.81 Band (50000-65000 MHz) MN4775A-0071 0.94 Band (65000-67000 MHz) MN4775A-0071 0.90 Band (67000-70000 MHz) MN4775A-0071 1.32 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 77: Me7848A-0270 1550 Nm Test Records

    MN4775A-0070 0.01 Band (05000-20000 MHz) MN4775A-0070 0.01 Band (20000-38000 MHz) MN4775A-0070 0.01 Band (38000-50000 MHz) MN4775A-0070 0.01 Band (50000-65000 MHz) MN4775A-0070 0.01 Band (65000-67000 MHz) MN4775A-0070 0.01 Band (67000-70000 MHz) MN4775A-0070 0.02 VectorStar ME7848A Series MM PN: 10410-00778 Rev. A...
  • Page 78 MN4775A-0070 0.46 Band (05000-20000 MHz) MN4775A-0070 0.46 Band (20000-38000 MHz) MN4775A-0070 0.58 Band (38000-50000 MHz) MN4775A-0070 0.81 Band (50000-65000 MHz) MN4775A-0070 0.94 Band (65000-67000 MHz) MN4775A-0070 0.90 Band (67000-70000 MHz) MN4775A-0070 1.32 PN: 10410-00778 Rev. A VectorStar ME7848A Series MM...
  • Page 80 Anritsu Company Anritsu utilizes recycled paper and environmentally conscious inks and toner. 490 Jarvis Drive Morgan Hill, CA 95037-2809...