Invacare® Carroll CS Series™
Effect
Actuator not working.
Actuator not working.
Bed ends to not fit properly.
58
Possible cause
Faulty actuator.
Faulty hand control pendant.
Bent brackets.
Verification
Disconnect power supply,
swap cable from unresponsive
actuator with a functional
actuator. Reconnect power
and test functions. Faulty
actuator will not work
regardless of the connection
port.
Pendant cable is connected,
power is on, lock-out
is off and hand control
pendant does not function.
Disconnect power supply
then replace hand pendant
with functioning unit (from
another bed, if available).
Reconnect power and test.
Visually ensure the brackets
are parallel to the frame.
Corrective action
Contact Invacare Continuing
Care Service Department
1-800-668-2337.
Contact Invacare Continuing
Care Service Department
1-800-668-2337.
Carefully straighten minor
deformations. Call Service
Department to replace
significantly bent brackets,
1-800-668-2337.
1153410-Q