4 Search methods
4.1 GPA search methods
There must be a probe connected to the electronic unit to use the GPA search
method.
4.1.1 Search with the universal probe
Switching the electronic unit on via On-/Off-button.
Select your language.
Select "GPA System" as search method.
The electronic unit should face up while using the probe
Optionally the tablet computer can be connected to the electronic unit via
Bluetooth. (See 5.2)
Place the universal probe so that one of the square ends shows in the
direction you want to search and check both, calibration value and sensor
value..
Turn the probe 180° so that the opposite square now face forward and check
the two values again.
Choose the side with values closer to 2,5 as forward direction (as shown in the
following figures).
The distance between the probe and ground should not vary a lot.
Avoid hectic movements and try to walk in straight lines.
Fig.: 14 - Universal probe
4.2 GOLD ID-XL search methods
There must be a coil connected to the electronic unit to use the GOLD ID-XL search
method.
KTS-Electronic GmbH & Co. KG – GPA 500 XS
Fig.: 15 - Complete universal probe
11