Setup; General Information; Adjustment Elements - Vega VEGAVIB 63 Operating Instructions Manual

Vibrating level switch with tube extension for granuled bulk solids
Hide thumbs Also See for VEGAVIB 63:
Table of Contents

Advertisement

Function/Configuration
Electronics and connec-
tion compartment
Adjustment of the density
range (1)
VEGAVIB 63 • Transistor (NPN/PNP)
6

Setup

6.1

General information

The figures in brackets refer to the following illustrations.
On the electronics module you will find the following display and

adjustment elements:

Potentiometer for adjustment of the density range (1)
DIL switch for mode adjustment - min./max. (2)
Signal lamp (5)
Note:
As a rule, always set the mode with mode switch (2) before starting
the setup of VEGAVIB 63. The switching output will change if you set
the mode switch (2) afterwards. This could possibly trigger other con-
nected instruments or devices.
6.2
Adjustment elements
5
4
Fig. 13: Electronics and connection compartment - transistor output
1
Potentiometer for adjustment of the density range
2
DIL switch for mode adjustment
3
Ground terminal
4
Connection terminals
5
Control lamp
With the potentiometer you can adapt the switching point to the solid.
It is already preset and must only be modified in special cases.
As a default setting, the potentiometer of VEGAVIB 63 is set to the
complete right position (> 0.1 g/cm³ or 0.0036 lbs/in³). In very light
solids you have to turn the potentiometer to the complete left posi-
tion (0.02 ... 0.1 g/cm³ or 0.0007 ... 0.0036 lbs/in³). By doing this,
6 Setup
1
2
3
19