Calibration Counter / Number Of Adjustments; Adjustment Mode And Displaying The Calibration Counter Contents; Adjustment Mode And Displaying The Calibration Counter Contents [284|285]; Placing Adjustment Weights On The Scale - Seca 374 Series Service Manual

Babyscale with nec ii electronic and radio module
Hide thumbs Also See for 374 Series:
Table of Contents

Advertisement

Maintenance
seca

2.1.2 Calibration Counter / Number of Adjustments

[374|376|378|703|704|284|285|954|957|959|963]
Scales fitted with the new, modular electronics are equipped with a calibration
counter allowing a software adjustment to be carried out in accordance with the
requirements for verified scales. Each completed adjustment procedure is
registered by the calibration counter, i.e. the number is automatically
incremented by 1. Non-approved scales are also fitted with a counter which
counts the number of adjustments.
2.1.3 Adjustment Mode and Displaying the Calibration Counter
Contents [374|376|378|703|704|954|957|959|963]
To begin readjustment or to display the calibration counter contents, start the
scale while pressing another key at the same time. The contents of the
calibration counter (i.e. the number of adjustment procedures carried out so far)
are then displayed for 18 seconds (flashing).
While the calibration count is displayed, a key must be kept pressed (for more
than 1.5 seconds) to switch the scale to adjustment mode.
2.1.4 Adjustment Mode and Displaying the Calibration Counter
Contents [284|285]
To begin readjustment or to display the calibration counter contents, start the
scale and press another key while the display shows seca. The contents of the
calibration counter (i.e. the number of adjustment procedures carried out so far)
are then displayed for 18 seconds (flashing).
While the calibration count is displayed, a key must be kept pressed (for more
than 1.5 seconds) to switch the scale to adjustment mode.

2.1.5 Placing Adjustment Weights on the Scale

[374|376|378|703|704|284|285|954|957|959|963]
Once you have switched to adjustment mode as described in section 0 or 2.1.4,
the display shows the text "CAL". By pressing any key except for the start key
for more than 1.5 seconds, the actual readjustment mode is activated. The
display reads "dec". The weight value currently measured then appears. The
scale is in decrementing state. To switch over from decrementing to
incrementing mode and vice versa press a key for more than 1.5 seconds with a
test weight placed on the scale. The display will read "dec" or "inc" accordingly.
The test weight is at least 25% of the max. load of the scale (we recommend
approx. 66-75%).
17.08.11/RLEU/ARI/MRE