Replacement instructions
seca 374/376

1 Opening the scale

(see Fig. 01)
 Open the battery compartment cover and remove the battery pack. Now unscrew
screw (B), push the electronics compartment cover (A) slightly to the right and remove
it.
Figure 01: Opening the scale

1.1 Opening the scale when removing load cells or the mains connection

(see Fig. 02)
 Additional step when replacing the load cell or the mains connection:
turn the scale round and place it on a soft mat. Remove the 4 screws (C, Fig. 02), then
carefully lift off the frame (D, Fig. 02).
Caution, the power connection socket is screwed to the frame. Take care not to tear
off the power connection cables.

2 Replacing

Please be sure to observe the sequence listed below when removing/fitting the parts!

2.1 Replacing the adjusting feet

(see Fig. 02)
 The adjusting feet (E) can be unscrewed and replaced by new ones without
disassembling the scale.
E
Figure 02: Replacing the adjusting feet
24.09.2010 /MRE
A
C
D
4
B
E
30-34-00-813