Configuration - Greiner Bio-One Sed Rate Timer 10/II User Manual

Esr analyzer, including software (g2m110bo, vers. 1.1),
Table of Contents

Advertisement

Connect the SRT10/II to the electricity supply by plugging the 12V power supply into the
intended power inlet on the SRT10/II. Only use the power supply provided.

3.3 Configuration

Before switching on the analyzer, please set-up the dip-switch placed in the rear panel
following the table below:
Switch
Function
1
Enable 15' analysis time.
2
Enable 60' analysis time (works only if switch 1 = off)
3
Enable additional result formats (works only if switch 1 = off)
4
Enable temperature compensation
5
Enable the printer
6
Enable sedimentation curve graph
7
Enable the internal fan
8
Enable power supply on printer connector ( DPT100 printer )
WARNING!: Do not switch on power when using a different type of
printer, as serious damage could be caused.
To enable each function be sure that the corresponding switch is in the ON position. For
the analysis time and the results formats the following configurations are possible:
SW1
On
Off
Off
Off
Off
(where 'x' = don't care, has no effect)
Example:
to configure instrument for 15 and 30 minutes analysis time, results corresponding to 1
hour Westergren, with temperature compensation, connected to the thermal printer (Item
number 836582), switches are placed: (Default Setting)
SRT 10/II
SW2
SW3
x
x
On
On
On
Off
Off
On
Off
Off
Analysis
Results format(s)
time
15'
mm/1hr
60'
mm/2hr & mm/1hr & mm/30'
60'
mm/2hr
30'
mm/1hr & mm/30'
30'
mm/1hr
USER MANUAL
page 9 of 37