3. INSTALLATION

3.1 Delivery contents
Check contents of carton using this list
• 1x SRT 10/II
• 1x 12-volt power supply
• 1x Power cable with Schuko plug
• 1x Control unit (contains three ESR Tubes with different long plastic sticks)
• 1x User Manual on an USB Stick
• 1x Dust cover
1x Certificate of Quality
Optional – not included in the packaging
836582
Thermal printer for SRT 10/II, SRS 20/II and SRS 100/II
836595
ACCUSED PLUS QC Control Blood for SRS

3.2 Positioning of the analyzer

The SRT 10/II must not be placed near centrifuges, oscillating agitators or other vibrating
instruments that might cause movement of the bench. Please keep in mind that ESR is very
sensitive to vibrations that could cause a false increase of results. The bench must be flat
and level. Direct light on the instrument and sudden changes of temperature should be
avoided. Keep a free area of at least 15 cm around the instrument to allow instrument cooling
by the internal back panel fan. Connect the instrument power supply only of the type
described in this User's Manual. Use the mains plug to disconnect the apparatus from the
mains supply. The mains plug must be accessible at all times.
Do not use this device in close proximity to sources of strong electromagnetic
radiation (e.g. unshielded intentional RF sources), as these may interfere with
proper operation.
Therefore the electromagnetic environment should be evaluated prior to operation
of the device.
The instrument is not intended to be integrated into a system.
Do not replace the detachable power cord by an inadequately rated power cord.
SRT 10/II
USER MANUAL
page 8 of 37