Greiner Bio-One Sed Rate Timer 10/II User Manual page 10

Esr analyzer, including software (g2m110bo, vers. 1.1),
Table of Contents

Advertisement

15 minutes analysis time
Function
Switch
1
Enable 15' analysis time.
2
Enable 60' analysis time (works only if switch 1 = off)
3
Enable additional result formats (works only if switch 1 = off)
4
Enable temperature compensation
5
Enable the printer
6
Enable sedimentation curve graph
7
Enable the internal fan
8
Enable power supply on printer connector (DPT100 printer)
(where 'x' = don't care, has no effect)
30 minutes analysis time
N° Switch
Function
1
Enable 15' analysis time.
2
Enable 60' analysis time (works only if switch 1 = off)
3
Enable additional result formats (works only if switch 1 = off)
4
Enable temperature compensation
5
Enable the printer
6
Enable sedimentation curve graph
7
Enable the internal fan
8
Enable power supply on printer connector (DPT100 printer)
SRT 10/II
USER MANUAL
Position
ON
x
x
ON
ON
OFF
ON
ON
Position
OFF
OFF
OFF
ON
ON
OFF
ON
ON
page 10 of 37

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Srt10/ii

Table of Contents