Greiner Bio-One Sed Rate Timer 10/II User Manual page 12

Esr analyzer, including software (g2m110bo, vers. 1.1),
Table of Contents

Advertisement

If the printer is connected, on the display appears this message:
Print curve ON...
Printer OK...
If the printer is not enabled the following message appears:
Printer OFF...
if the temperature compensation is active, the following message appears on the display:
18°C temp. of reference
25.0°C of internal temp.
Instrument starts the auto test for mechanical and electronic control
self-test...
Several seconds after, the display will indicate:
self-test OK...
Now the instrument is ready for the analysis. On the display appears this message:
. . . . . . . . . . results: 1h
temp.25.0°C
The top row will display the test time remaining, pre-indication or results as explained in
section 'System Description'. When multiple results formats are enabled, the display will
alternate between the selected formats.
SRT 10/II
USER MANUAL
page 12 of 37