6. TEMPERATURE COMPENSATION
6.1 Results correction to 18°C
The results achieved are correlated to the method of reference, considering the room -
temperature. SRT 10/II, measures constantly the inner temperature; further reconverts the
values in according to the Manley table (1) shown below, at the temperature of 18 degrees.
Therefore, this instrument guarantees a better reproducibility instead of instruments which
perform results without temperature compensation.
(1)
correct
values
____________________________________________________________
18°C
5
10
20
30
40
50
60
70
80
90
100
SRT 10/II converts the results to 18 degrees according to the table if room temperature is in
the range 15 – 32°C. For lower or higher room temperatures the instrument convert
temperature in this way: 15°C for lower and 32°C for higher temperature.
6.2 Result
Each laboratory should validate and evaluate its ESR reference values.
SRT 10/II
Manley table
------------ analysis temperatures -----------------------
15°C
18°C
4
5
9
10
18
20
27
30
36
40
46
50
55
60
63
70
72
80
81
90
90
100
USER MANUAL
20°C
25°C
5
6
10
12
21
25
31
37
42
49
52
60
62
71
72
82
82
93
93
103
103
114
30°C
8
16
31
45
58
71
82
93
104
114
125