266

Manual start of the automatic calibration

Automatic calibration
Manual start of the automatic calibration
Manual stop of the automatic calibration
Validation
AO2000 CONTINUOUS GAS ANALYZERS
NOTES
Calibration should only be started after the warm-up phase.
For the Caldos25, Caldos27, Magnos206, Magnos28 and Magnos27 analyzer
modules, it is not admissible to perform only a span calibration. A zero cali-
bration must always precede a span calibration.
Automatic calibration can be performed
as zero calibration alone or
as span calibration alone or
jointly as zero and span calibration.
1
Select the Automatic calibration menu:
MENU → Calibrate → Automatic calibration
2
Only zero-point calibration: ZERO AUTOCAL
Only span calibration: SPAN AUTOCAL
Common Zero and Span Calibration: ZERO & SPAN AUTOCAL
The user can end the automatic calibration process by pressing the STOP
softkey.
When automatic calibration is stopped, the analyzer module is in an indefi-
nite state. For example, the zero-point calibration may have been completed
and calculated, but the end-point calibration has not yet been carried out.
For this reason, automatic calibration will have to be restarted and allowed
to run to completion after any cancelation of automatic calibration.
The procedure described above applies in the same way if the analyzer mod-
ule is in validation (see page 230) operating mode.
|
OI/AO2000-EN REV. B