Accuracy To Ysi Within Cgm Glucose Ranges - Abbott FreeStyle Libre 2 User Manual

Flash glucose monitoring system
Hide thumbs Also See for FreeStyle Libre 2:
Table of Contents

Advertisement

Table 1b: Accuracy to YSI within CGM Glucose Ranges (Adult; n=144)
CGM
Number
Glucose
of CGM-
Level
Reference
(mg/dL)
Pairs
<54
518
54-69
3012
70-180
7785
181-250
3037
>250
4383
System range is 40-400 mg/dL.
133
Percent
Percent
Within
Within
±15 mg/dL
±20 mg/dL
85.9
93.8
89.5
94.2
Percent
Percent
Percent
Within
Within
Within
±40 mg/dL
±15%
±20%
99.4
99.1
76.5
89.1
94.0
Percent
Mean
Within
bias
±40%
(mg/dL)
-6.4
-3.3
86.6
99.2
-4.8
95.0
99.9
-10.1
97.9
100.0
-6.3
MARD
(%)
13.8
10.8
10.6
7.8
6.1

Hide quick links:

Advertisement

Table of Contents
loading