IMPORTANT: (cont.)
Inaccurate test strip results may occur in severely hypotensive
individuals or patients in shock.
Inaccurate test strip results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without ketosis.
Observe caution when using around children. Small parts may
constitute a choking hazard.
You should clean and disinfect the Reader once per week. The
Reader should also be cleaned and disinfected prior to being
handled by any person providing testing assistance to the user.
The Reader is for use by a single person. It must not be used on
more than one person including other family members due to the
risk of spreading infection. All parts of the Reader are considered
biohazardous and can potentially transmit infectious diseases, even
after performing the cleaning and disinfection procedure.
Use the Reader's built-in meter within the test strip operating
temperature range (59˚F – 104˚F) or you will see Error Message E-1.
Use a test strip immediately after removing from its foil packet.
Only use a test strip once.