Electrical Connections - Aalborg DPM Series Operating Manual

Intelligent digital mass flow meters
Table of Contents

Advertisement

3.

ELECTRICAL CONNECTIONS

DPMisequippedwithan 8 pin-MiniDINpower,analog/relayoutput,
communication interface connector. Table I explains the pin designations. See
Figure 1 for a Pin Diagram.
TABLE I: 8-PIN DESIGNATIONS AND NOTES
PIN
FUNCTION
1
Solid State SPST Relay NO
(normally open) contact #1
2
Solid State SPST Relay NO
(normally open) contact #2
3
RS-232 RX / RS-485 (–)
Communication Interface input
4
Analog (0-5Vdc,0-10Vdc,4-20
mA) Output reference (-)
5
RS-232 TX / RS-485(+)
Communication Interface input
6
Analog (0-5Vdc, 0-10Vdc or
4-20 mA) Output (+)
7
Power supply, positive (+)
8
Powersupply,common (-)
RS-232 SignalGround
CAUTION:
FIGURE 1: DPM 8-PIN Mini-DIN CONNECTOR CONFIGURATION
NOTE
Do not exceed SSR maximum voltage 48
AC peak/DC and maximum load current 400
mA.
Also accessible via Audio jack connector
(see Figures 2 & 25)
Common (return) for pin 6 (0-5Vdc or
0-10 Vdc or 4-20 mA)
Also accessible via Audio jack connector
(see Figures 2 & 25).
Output. Do not apply external voltage
or any current source. Be sure to observe
recommended load impedance.
Power input 9 – 26 Vdc.
Power input and RS-232 communication
common.
4-20 mA analog output requires at least 12 Vdc power.
6

Hide quick links:

Advertisement

Table of Contents
loading