UNPACKING THE GFM MASS FLOW METER Inspect Package for External Damage Your GFM Mass Flow Meter was carefully packed in a sturdy cardboard carton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package contact the shipping company immediately.
CAUTION: GFM TRANSDUCERS SHOULD NOT BE USED FOR MONITORING OXYGEN GAS UNLESS SPECIFICALLY CLEANED AND PREPARED FOR SUCH APPLICATION. For more information, contact your distributor or Aalborg 7 . Attitude limit of the Mass Flow Meter is ±15 from calibration position (standard cal- ibration is in horizontal position).
Page 6
FIGURE 2.a - 9-PIN “D” CONNECTOR PINOUTS FOR GFM TRANSDUCER. FUNCTION Remote LCD display signal 0 to 5 VDC output indication 0 to 5 VDC common Power supply, positive Power supply, common Remote LCD display reference (unassigned) 4 to 20 mA output indication 4 to 20 mA common IMPORTANT NOTES: In general, “D”...
2.3.1 Remote LCD Readouts GFM Mass Flow Meters are available with optional remote reading LCD displays supplied with a three foot long wire to accommodate most applications. This con- figuration includes the upper block element which serves as the LCD readout mounting.
ATTITUDE SENSITIVITY: Incremental deviation of up to 1% full scale from stated accura- cy, after re-zeroing. OUTPUT SIGNALS: Linear 0-5 VDC (1000 Ω minimum load impedance) and 4-20 mA (0-500 loop Ω resistance); 20 mV peak to peak max noise. Contact your distributor or Aalborg 7 for optional RS232 or IEEE488 interfaces.
GFM17S/37S/47S/57S/67S/77S: 316 stainless steel with VITON7 O-rings seals; BUNA7, NEOPRENE7 or KALREZ7 O-rings are optional. Aalborg 7 makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the different models.
Page 10
FLOW RANGES TABLE I GFM 17 TABLE II GFM 37 LOW FLOW MEDIUM FLOW MASS FLOW METER* MASS FLOW METER* scc/min [N CODE CODE std liters/min [N 0 to 10 0 to 15 0 to 20 0 to 50 0 to 100...
The full scale range and gas for which your meter has been calibrated are shown on the flow transducers front label. For optional RS232 or IEEE488 interfaces please contact your distributor or Aalborg 7 .
Swamping Condition If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter is taking place, a condition known as “swamping” may occur. Readings of a “swamped” meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range.
6.2.2 GFM 17 models Unscrew the inlet compression fitting of meter. Note that the Restrictor Flow Element (RFE) is connected to the inlet fitting. Carefully disassemble the RFE from the inlet connection. The 50 micron filter screen will now become visible. Push the screen out through the inlet fitting. Clean or replace each of the removed parts as necessary.
Remove the block and connected screens. Clean each of the removed parts as necessary. If alcohol is used for cleaning, allow time for drying. Inspect the flow path inside the transducer for any visible signs of contaminants. If necessary, flush the flow path through with alcohol. Thoroughly dry the flow paths by flowing clean dry gas through.
FIGURE 7.a - CALIBRATION POTENTIOMETER AND JUMPER LOCATIONS Calibration of GFM Mass Flow Meters All adjustments in this section are made from the outside of the meter, there is no need to disassemble any part of the instrument. GFM Mass Flow Meters may be field recalibrated/checked for the same range they were originally factory calibrated for.
7.2.3 SPAN Adjustment Reconnect the gas source. Using the flow regulator, adjust the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. If the devia- tion is less than ±10% of full scale reading, correct the SPAN potentiometer [R33] setting by using the insulated screwdriver through the access window, to eliminate any deviation.
7.3.3 25% Flow Adjustment Reconnect the gas source. Using the flow regulator, adjust the flow rate to 25% of full scale flow. Check the flow rate indicated against the flow calibrator. Adjust the set- ting for potentiometer [R33] by using the insulated screwdriver through the access window, until the output of the flow meter reads 1.25VDC ±63mV (or 8mA ±0.25mA).
7.4.2 Adjust Scaling Using a digital multimeter connected to either the 0 to 5 VDC or 4 to 20 mA sig- nal at the 9-pin “D” connector, set the flow rate on the GFM to full scale flow (5 VDC or 20mA). Maintain full scale flow, and adjust the potentiometer [R3] on the LCD printed circuit board to desired full scale flow reading.
Troubleshooting Guide INDICATION LIKELY REASON REMEDY lack of reading or output power supply off check connection of power supply fuse blown disconnect GFM transducer from power supply; remove the shorting condition or check polarities; fuse resets automatically filter screen flush clean or disassemble obstructed at inlet to remove impediments or replace...
Page 20
INDICATION LIKELY REASON REMEDY No zero reading after 15 Embedded temperature has Readjust ZERO minute warm up time and no been changed. potentiometer R34 through flow condition. the access hole (see page 12 for details). No zero reading after 15 Power supply voltage is Measure voltage on pins 4 minute warm up time and no...
Page 21
INDICATION LIKELY REASON REMEDY LCD Display reading does LCD Display is adjusted for Readjust LCD Display not correspond the correct wrong flow range or scaling for required full scale flow range according analog engineering units. flow (see 7.4 on page 14). output 0-5 Vdc signal.
Page 22
LIKELY REASON INDICATION REMEDY LCD Display reading is Sensor under swamping Lower the flow through GFM conditions (flow is more above maximum flow range within calibrated range or than 10% above maximum and output voltage 0-5 Vdc shut down the flow com- flow rate for particular signal is more than 5.5 Vdc pletely.
INDICATION LIKELY REASON REMEDY GFM is disconnected from Sensor or PC board is Return GFM to factory for defective. the source of the gas (no repair. flow conditions) but LCD Display reading is fluctuating in wide range. Output voltage 0-5 Vdc signal also fluctuating.
CALIBRATION CONVERSIONS FROM REFERENCE GASES The calibration conversion incorporates the K factor. The K factor is derived from gas density and coefficient of specific heat. For diatomic gases: d X C where d = gas density (gram/liter) = coefficient of specific heat (cal/gram) Note in the above relationship that d and Cp are usually chosen at the same con- ditions (standard, normal or other).
APPENDIX 2 GAS FACTOR TABLE (“K” FACTORS) CAUTION: K-Factors at best are only an approximation. K factors should not be used in applications that require accuracy better than +/- 5 to 10%. K FACTOR Density ACTUAL GAS Relative to N [Cal/g] [g/I] Acetylene C...
Page 27
K FACTOR Density ACTUAL GAS Relative to N [Cal/g] [g/I] Deuterium D 1.00 1.722 1.799 .4357 .508 1.235 Diborane B .1947 9.362 Dibromodifluoromethane CBr Dichlorodifluoromethane (Freon-12) CCl .3538 .1432 5.395 .4252 .140 4.592 Dichlofluoromethane (Freon-21) CHCl Dichloromethylsilane (CH SiCl .2522 .1882 5.758 .4044...
Page 28
K FACTOR Density ACTUAL GAS Relative to N [Cal/g] [g/I] 1.000 .0861 3.610 Hydrogen Bromide HBr 1.000 .1912 1.627 Hydrogen Chloride HCl .764 .3171 1.206 Hydrogen Cyanide HCN .9998 .3479 .893 Hydrogen Fluoride HF .9987 .0545 5.707 Hydrogen Iodide HI Hydrogen Selenide H .7893 .1025...
Page 29
K FACTOR Density ACTUAL GAS Relative to N [Cal/g] [g/I] Phosphorous Oxychloride POCl .1324 6.843 Phosphorous Pentafluoride PH .3021 .1610 5.620 Phosphorous Trichloride PCl .1250 6.127 Propane C .399 1.967 Propylene C .366 1.877 Silane SiH .5982 .3189 1.433 Silicon Tetrachloride SiCl .284 .1270 7.580...
This product is not intended to be used in life support applications! TRADEMARKS Aalborg Neoprene ® ® -is a registered trademark of Aalborg Instruments & Controls. -is a registered trademark of DuPont. Buna Viton ® ® -is a registered trademark of DuPont Dow Elastometers.