Installation Note
Part Number N9020-90240
Printed in USA August 2014

Advertisement

loading

Summary of Contents for Keysight 9020A

  • Page 1 Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement busi- ness has become Keysight Technologies. For more in- formation, go to www.keysight.com. Installation Note Part Number N9020-90240 Printed in USA August 2014...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
  • Page 3 Option HLC, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Option HLC, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Products Affected: N9020A, MXA Signal Analyzer ≥ MY52330000 Serial Numbers: ≥ SG52330000 ≥...
  • Page 4 Option HLC, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Software and test equipment is required for making adjustments and for performance verification testing. Information on how to obtain this software can be found at: www.agilent.com/find/calibrationsoftware NOTE Instrument software revision A.13.15 or later is required to install this upgrade.
  • Page 5 Option HLC, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Installation Kit Parts List Quantity Description Agilent Part Number A25 Wideband Analog IF N9020-60044 A26 Wideband Digital IF N9020-60257 Opt HLB/HLC Cable Kit with Wire Markers N9020-60153 (includes cables W51 through W56, listed below) Cable Assembly, Flat Flexible 80-conductor 3-in-LG (W44)
  • Page 6 Option HLC, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix ≥ MY/SG/US5233 Tools Required • T-10 TORX Driver • T-20 TORX Driver • Agilent Calibration and Adjustment Software, N7814A (revision E.11.00 or later) • Test equipment and computer supported by the X-Series Performance Tests and Adjustment Software •...
  • Page 7: Installation Procedure

    Installation Procedure Installation Procedure Analyzer Information 1. Connect a power cord to the analyzer and turn on the analyzer. 2. After the analyzer has completed turning on, press System Show System . Make note of the following information from the Show System screen: Product Number ________________ Serial Number ________________ Instrument S/W Revision ________________...
  • Page 8: Update Instrument Software

    Installation Procedure Update Instrument Software Updating the instrument software and installing the necessary licenses before installing the new hardware will help ensure that the hardware installation was successful. NOTE The minimum instrument software version is A.13.15. Go to the following website and determine whether or not the analyzer has the latest instrument software already installed: http://www.agilent.com/find/xseries_software If the analyzer does not have the latest instrument software already installed, download and install the...
  • Page 9 Installation Procedure Licensing the New Options NOTE If N9020AK-BU4, Analysis Bandwidth Upgrade, 85 MHz to 125 MHz or N9020AK-BU5, Analysis Bandwidth Upgrade, 85 MHz to 160 MHz will also be installed, the licenses provided by these kits should be installed at this time. 1.
  • Page 10 Installation Procedure 7. The signal analyzer will automatically consume the License File. (This may take a few minutes) When the License File is consumed the Agilent License Manager will display a “Successful License Installation” message as shown in Figure 2. Since the license file contains multiple licenses, multiple “Successful License Installation”...
  • Page 11: Analyzer Disassembly

    Installation Procedure Analyzer Disassembly CAUTION If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. NOTE If the analyzer has Option PRC, Portable Configuration, refer to the “Portable...
  • Page 12 Installation Procedure Figure 3 Standard Instrument Outer Case Removal 5. Proceed to the Front Frame Assembly Removal section to remove the front frame. Installation Note N9020-90240...
  • Page 13 Installation Procedure Portable Instrument (Option PRC) NOTE Make sure any adapters on the front panel are removed. 1. Disconnect the instrument from ac power. 2. Refer to Figure 4. Using the T-20 driver, remove the four screws (two on each side) (1) that hold the bail handle (2) to the front frame.
  • Page 14 Installation Procedure 4. Refer to Figure 5. Using the T-20 driver, remove the four screws including washers (1) that hold the rear bumpers (2) in place. 5. Pull the instrument cover (3) off towards the rear of the instrument. Figure 5 Option PRC Instrument Outer Case Removal Installation Note N9020-90240...
  • Page 15 Installation Procedure Front Frame Assembly Removal 1. Refer to Figure 6. Using the T-10 driver, remove the eight screws (1), four on each side, to detach the front frame from the chassis. Figure 6 Front Frame Removal Installation Note N9020-90240...
  • Page 16 Installation Procedure 2. Refer to Figure 7. Pull the front frame carefully away from the chassis. Remove the ribbon cable W1 from the A8 Motherboard. NOTE W1 may have locking springs on each side. Depress the spring on each side of the connector to remove from the motherboard.
  • Page 17 Installation Procedure Top Brace and Reference Bracket 1. Refer to Figure 8. To remove the top brace (1) use the T-10 driver to remove the eight panhead screws (3) (0515-0372), four on each side, attaching the brace to the chassis. Also remove the ten flathead screws (4), (7) (0515-1227) attaching the top brace to the boards.
  • Page 18 Installation Procedure RF Bracket Removal 1. Refer to Figure 9. Remove the RF bracket (1) by removing the remaining seven screws (2) using the T-10 driver. Figure 9 RF Bracket Removal Installation Note N9020-90240...
  • Page 19 Installation Procedure Removing Cables 1. Refer to Figure 10. Remove the wire cable hold down on the right side of the chassis to free the gray coax cables going to the A16 Reference Assembly. 2. Remove cable ties from bundle of gray coax cables. Figure 10 Cables Under Cable Hold Down Installation Note...
  • Page 20 Installation Procedure 3. Refer to Figure 11. Remove cable W50 connecting A3 Digital IF J17 to A16 Reference Assembly J726. Note cable routing from Digital IF J17 through the opening in the side panel. This cable will not be reused. 4.
  • Page 21 Installation Procedure Replace A6 Power Supply Assembly 1. Refer to Figure 12. Using a 5/16-inch wrench, disconnect semi-rigid cables W5, W6, and W19 from the A16 Reference Assembly (1). 2. Disconnect flexible coax cables W14, W20, and W23 from the A16 Reference Assembly (1). 3.
  • Page 22 Installation Procedure 4. Refer to Figure 13. Remove the six screws (1) attaching the CPU assembly to the chassis. 5. Remove the CPU assembly from the chassis by pulling straight out the back. Use the two ejectors to pull the CPU assembly out from the chassis. Figure 13 Current CPU Assembly - Rear Panel View Installation Note...
  • Page 23 Installation Procedure 6. Refer to Figure 14. Remove the three screws (1) attaching the power supply (2) to the power supply bracket (3). Keep these screws; they will be used later. Figure 14 Power Supply Assembly Removal 7. Remove the power supply assembly from the chassis by pulling straight out the back. 8.
  • Page 24 Installation Procedure 11. Replace the six screws (1) that attach the CPU assembly to the chassis. Torque to 9 inch-pounds. 12. Slide the reference assembly into the slot at the rear of the instrument and push on the assembly to mate the connectors to the Midplane assembly.
  • Page 25 Installation Procedure Installing Boards and Cables 1. Refer to Figure 16. Locate ribbon cable 8121-1854 (W44) and the A26 Wideband Digital IF in the kit. Connect W44 to the ribbon cable connector on the A26 Wideband Digital IF. 2. Install the A26 Wideband Digital IF into slot 4 by pressing the ejector. Lift the free end of the ribbon cable up as shown in Figure Figure 16...
  • Page 26 Installation Procedure 6. Refer to Figure 17. Locate W55 in the Opt HLB/HLC Cable Kit. This cable will have part number 8121-2290 and will have ends labeled “805” and “17”. Connect the end labeled “17” to A3J17 on the A3 DIF Assembly. Route the cable through cutout in the right side chassis near A10 Attenuator B and connect the other end to A25J805 on the A25 Wideband Analog IF Assembly.
  • Page 27 Installation Procedure 8. Refer to Figure 18. Locate W56 in the Opt HLB/HLC Cable Kit. This cable will have part number 8121-2291 and will have ends labeled “806” and “726”. Connect the end labeled “806” to A25J806 on the A25 Wideband Analog IF Assembly. Route the cable along the right side chassis behind attenuators A9 and A10 and connect the other end to A16J726 on the A16 Reference Assembly.
  • Page 28 Installation Procedure 10. Locate W53 in the Opt HLB/HLC Cable Kit. This cable will have part number 8121-2292 and will have ends labeled “901” and “101”. Connect the end labeled “901” to A15J901 on the A15 Enhanced Front End Control Assembly. Route the cable along the right side chassis and connect the other end to A25J101 on the A25 Wideband Analog IF Assembly.
  • Page 29 Installation Procedure Installation of N9020AK-HL9 or N9020AK-HLA Simultaneously with N9020AK-HLC NOTE If you are also installing either N9020AK-HL9, 40 MHz Analysis Bandwidth (B40) for Opt 503 or N9020AK-HLA, 40 MHz Analysis Bandwidth (B40) and Microwave Preselector Bypass (MPB) upgrade kit at the same time as the N9020AK-HLC kit, continue with the Dressing Cables procedure in the installation note for N9020AK-HL9 or N9020AK-HLA, as appropriate.
  • Page 30 Installation Procedure Reassembly 1. Refer to Figure 21. Reinstall the RF bracket (1) onto the chassis using seven screws (2). Torque to 9 inch-pounds. Figure 21 RF Bracket Replacement Installation Note N9020-90240...
  • Page 31 Installation Procedure 2. Refer to Figure 22. To replace the top brace (1), place it in the correct position and attach the appropriate screws. Torque to 9 inch-pounds. Figure 22 Top Brace and Reference Bracket Replacement 3. Locate the rear panel replacement kit, N9020-60149, included in the kit. Use the rear panel from this kit in the following steps.
  • Page 32 Installation Procedure 7. Refer to Figure 23. Place the rear panel (1) into position in the chassis. Replace the screws (2) to attach the rear panel to the chassis. Torque to 9 inch-pounds. Figure 23 Rear Panel Replacement Installation Note N9020-90240...
  • Page 33 Installation Procedure 8. Refer to Figure 24. Reattach the ribbon cable W1. Figure 24 Front Panel Cable Installation Note N9020-90240...
  • Page 34 Installation Procedure 9. Refer to Figure 25. Carefully position the Front Frame Assembly onto the chassis. Ensure no cables are crushed. Replace the eight screws (1), four on each side of the chassis. Torque to 9 inch pounds. Figure 25 Front Frame Replacement Installation Note N9020-90240...
  • Page 35 Installation Procedure Final Installation for Standard Instruments (Benchtop Configuration, Figure 1. Refer to Figure 3. Carefully slide the instrument cover back onto the instrument from the rear of the analyzer, making sure not to damage any internal cables. The seam on the cover should be on the bottom of the instrument.
  • Page 36 Installation Procedure Power Up and New Hardware Wizard 1. Connect a keyboard and mouse to the instrument. 2. Power on the instrument. 3. During the boot up process you may notice that the “Found New Hardware” bubble appears in the lower right screen, and a “Found New Hardware’...
  • Page 37 Installation Procedure 5. Enter administrator as the user name, and agilent4u as the password. Select OK. 6. The screen in Figure 27 appears. Select Yes, This time only. Click Next. Figure 27 Installation Note N9020-90240...
  • Page 38 Installation Procedure 7. The screen in Figure 28 appears. Ensure “Install the software automatically’ is selected and click Next. Figure 28 8. The wizard will install the required software. Once you see the “Completing the Found New Hardware Wizard” screen appear, click Finish. Installation Note N9020-90240...
  • Page 39: Update The Instrument Software

    Installation Procedure Update the Instrument Software NOTE Instrument software version A.13.15 or later is required. 1. Loading the latest instrument software is required to assure all FPGAs and drivers located on the newly installed hardware and in the base instrument are synchronized. Therefore, even if the instrument contains the latest revision of software, you must reinstall the software to assure proper operation.
  • Page 40 Installation Procedure Verify the Option 1. Turn the analyzer on and verify no errors are reported at power up through the power-on alignment of the analyzer. 2. Press Mode IQ Analyzer (Basic) 3. Press , or , if available). Mode Setup IF Path 85 MHz 125 MHz...
  • Page 41: Adjustments Required

    Utilities, Adjustments, and Performance Verification Tests Utilities, Adjustments, and Performance Verification Tests Calibration software and specified test equipment is required to perform the adjustments and can be used to automate the performance verification testing. Obtain Agilent X-Series Signal Analyzer Calibration Application SW, N7814A TME Calibration Application, version E.11.00 or later.
  • Page 42 Utilities, Adjustments, and Performance Verification Tests Installation Note N9020-90240...