Care Instructions

Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
Before use by a different individual: clean, disinfect, and check product for proper
functioning.
DO NOT:
- machine wash or dry the cushion or the Convex Insert
- immerse the cushion or the Convex Insert in water
- autoclave, steam clean, or expose cushion or the Convex Insert to temperatures in
excess of 150˚F (65˚C)
- use caustic, automatic dishwasher detergents
- use cleaning products containing petroleum or organic solvents such as acetone,
toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids, or adhesive removers
- expose the cushion or the Convex Insert to ozone gas or to ultraviolet light cleaning
methods
Before cleaning and disinfecting, remove the cover and the Convex Insert from the
Note:
cushion. Remove hook fastener pieces from the bottom of the cover.
Machine wash in cold water (30°C) with mild detergent,
To Clean the Cover:
gentle cycle. Tumble dry low.
To Disinfect the Cover:
using 1 part household liquid bleach per 9 parts water,
then rinse thoroughly. The cover can also be machine
washed in hot water (60°C) to disinfect. Tumble dry low.
Follow safety guidelines on bleach container.
To Clean the Cushion and the Convex Insert:
dampened with water and a household detergent. Rinse cloth and re-wipe. Allow
foam to air dry thoroughly before putting the cover on or replacing the Convex Insert.
Disinfectants are not effective on porous surfaces such as foam. If the cushion or the
Convex Insert becomes soiled or contaminated, DO NOT use with multiple individuals. Replace
soiled or contaminated foam components.
This symbol means, "Wipe with a damp cloth."
ROHO AirLITE Cushion Operation Manual (T22305)
Hand wash in warm water (40°C)
Remove the cover. Wipe gently with a cloth
+