LumaSense technologies IMPAC IGA315-K USB Operation Manual

LumaSense technologies IMPAC IGA315-K USB Operation Manual

Handpyrometer

Advertisement

Quick Links

上海麦兴仪器设备有限公司
www.imaxsun.com
IMPAC IGA315-K USB
Handpyrometer - Handpyrometer
Operation Manual - Betriebsanleitung
Proven Quality

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IMPAC IGA315-K USB and is the answer not in the manual?

Questions and answers

Summary of Contents for LumaSense technologies IMPAC IGA315-K USB

  • Page 1 上海麦兴仪器设备有限公司 www.imaxsun.com IMPAC IGA315-K USB Handpyrometer - Handpyrometer Operation Manual - Betriebsanleitung Proven Quality...
  • Page 2: Table Of Contents

    6.10. Factory settings 7. Software 7.1. Installation and how to use 8. Maintenance 8.1. Rechargeable batteries 8.2. Cleaning 9. lMPAC portable pyrometer Art.-No. 10. Spare parts 11. Installation USB Driver LumaSense Technologies Kleyerstr. 90 D - 6036 Frankfurt GERMANY www.Lumasenseinc.com impac@Lumasenseinc.com...
  • Page 3: Technical Data

    Heat protection cover Grey filter Mounting key for Greyfilter High temperature protective plate Hand strap Recharger Typ ASC 410 for internal recharging Software for IMPAC IGA315-K USB 3. Additional equipment Carrying case Art.-No.: 4 344 010 Infrared interface IRS-315 USB Battery charger type Voltcraft for external recharging 6 rechargable batteries 1. V ;...
  • Page 4: Operation

    Two step Strap connection compartment measuring button Ocular Recharging joint, 3.5 mm Tubering Grey filter Protective glass Rubber eye-pice Tightening ring Fixing for Handstrap x M6-thread for heat protection shield Pict. : User-interface of IMPAC IGA315-K USB right side view...
  • Page 5: Operation

    IMPAC Handpyrometer IGA315-K USB 5.2. Operation Before operating the Pyrometer IGA315-K USB the looking though lense is to be adjusted at a bright, white background (e.g. a sheet of paper) by turning the ocular until the circle-shaped measuring field is visible sharply at the relaxed eye. We recommend the using of the grey filter by screwing in the thread of the ocular at high measuring temperatures. Focusizing is done by the looking through lens. To adjust the distance loosen the tighting ring and move the tube by turning until the measured object is focusized sharply.
  • Page 6: Changing Rechargeable Batteries / External Recharging

    IMPAC Handpyrometer IGA315-K USB 5.4. Changing rechargeable batteries / external recharging To charge the rechargeable batteries the ribbed screw of the battery compartment must be loosened and the compartment cover must be removed. Now you can remove the rechargeable batteries. The batteries must inserted in the battery charger correctly as shown in the battery chargers manual.
  • Page 7: Clearing Of Storage

    IMPAC Handpyrometer IGA315-K USB 5.7. Clearing of storage To initiate clearing press simultaneously. Select Init. Display: Confirmation of clearing is done by pressing Display : after a few seconds Abortion of clearing is done by pressing before confirmation. Select 5.8. Data transmission to PC With the included software, it is easily possible to read out the measured data. This leads to the positioning pins of the included infrared interface in the positioning of support is an on pyrometer.
  • Page 8: Battery Test

    IMPAC Handpyrometer IGA315-K USB 5.9. Battery test Conf. The battery test will be confirmed by pressing and holding the key. Display at sufficient battery power:: If the battery power is beneeth 5.5 V it will be displayed as follows: The batteries should be recharged or changed. Attention: Never use not rechargable batteries !!
  • Page 9: Configuration

    IMPAC Handpyrometer IGA315-K USB 6. Configuration 6.1. Selection of values The selection of values can be done by pressing the key. The displayed value will be enlarged or reduced by one step. If one of the two keys is pressed longer than two seconds a quick-count mode wil be setted to allow the change of values more quickly.
  • Page 10: Minimum Temperature

    IMPAC Handpyrometer IGA315-K USB 6.4. Minimum temperature Display : The minimum temperature can be chosen by pressing between 600 °C (1112 °F) and 1600 °C (2912 °F). Notice: If the measured value is beneath this minimum temperature will not be stored! 6.5. Energy saving mode The energy saving mode will shutdown energy intensive processes (e.g. LED dispays). These status will be reached after not using the instrument for the presetted time (in se- conds). The instrument will not shut down completely and will be in measuring mode immi- deatly after pressing the measuring button.
  • Page 11: Measuring Mode

    IMPAC Handpyrometer IGA315-K USB 6.8. Measuring mode Display: The measuring mode can be selected by pressing . Possible modes are: (C): Continous measuring: The measured values will be displayed continously in the display as long as the measuring key is pressed at the first step. The actually measuring value will be written to the storage if the measuring key is pressed at the second step. (P): Peak picker mode: The measured values are computed continously to an actual maximum value.
  • Page 12: Software

    IMPAC Handpyrometer IGA315-K USB 7. Software This software has been designed to pick out the pyrometer readings comfortably. Readings can be shown easily and clear either in a table or further to analyze the data you get the opportunity to transfer it into a MS-Excel file. 7.1. Installation and how to use System requirements: Standard PC with Win 98 or higher...
  • Page 13: Lmpac Portable Pyrometer Art.-No

    IMPAC Handpyrometer IGA315-K USB 9. lMPAC portable pyrometer Art.-No. Item: Art. - No.: IMPAC IGA315-K USB, portable pyr.f. coking plants, 600 … 1600°C, focus.optics, Basic config. *) 4 327 130 IMPAC IGA315-K USB, portable pyr.f. coking plants, 600 … 1600°C, focus.optics, Full config.**) 4 327 140 *) Scope of delivery Basic configuration: Accumulator set (6 pcs, 1.2V,2100 mAh, NiMH), Heat Protection Jacket, Neutral Gray Filter in Holder, Tubular Key for Gray Filter, High Temperature Protection Glass, Wrist Strap, Charging Unit ASC 410 (internal charging), Operation Manual **) Scope of delivery Full configuration: see Basic configuration plus: Case with Foam Inlay, Infrared Interface IRS 315 USB with Software, Power Supply, Charging Unit Voltcraft (external charging), Strap with Carabiner 10. Spare parts Item: Art.
  • Page 14: Installation Usb Driver

    IMPAC Handpyrometer IGA315-K USB 11. Installation USB Driver The installation consists of two steps. First of all is the installation as USB serial converter. The please install the USB serial port. Please insert the CD in your CD-ROM drive. As soon as the device is connected the hardware assistant will detect it automatically.
  • Page 15 IMPAC Handpyrometer IGA315-K USB Please choose in the following window the appropriate source (your CD-ROM drive) with the driver. The software will be searched under the given source path and than will be installed. You have now finished the installation of USB serial converter. Installation of USB serial port will be done analogue to USB serial converter (please see previous figures).